Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry
Table 2
Intraocular pressure readings with Perkins, air and Diaton tonometers, and central and superior pachymetries before and after LASIK.
Parameter
Presurgery
Postsurgery
Presurgery − postsurgery
value
IOP Perkins (mmHg) (mean (SD))
14.02 ± 2.25
11.85 ± 2.08
2.16 ± 2.47
IOP air tonometry (mmHg) (mean (SD))
14.92 ± 2.98
10.86 ± 1.71
4.05 ± 2.72
IOP Diaton (mmHg) (mean (SD))
14.18 ± 2.70
14.38 ± 3.50
−0.20 ± 2.73
0.590
Central pachymetry (µm) (mean (SD))
572.59 ± 45.41
475.91 ± 55.95
96.67 ± 44.83
Superior pachymetry (µm) (mean (SD))
658.62 ± 31.69
645.18 ± 29.41
13.44 ± 29.30
0.080
Presurgery versus postsurgery. . Student’s paired -test.