Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes

Table 1

Intraobserver repeatability of the corneal dynamic Scheimpflug analyzer Corvis ST in measuring central corneal thickness ().

OperatorMean (μm) ± SD (μm)TRT (μm)CoV (%)ICC (95% CI)

1st535.9 ± 27.04.813.00.870.971
(0.958 to 0.980)
2nd537.4 ± 27.64.713.00.870.972
(0.960 to 0.981)

SD = standard deviation, = within-subject standard deviation, TRT = test-retest repeatability (2.77), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.