Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes

Table 2

Interobserver reproducibility of central corneal thickness readings using average (from average of 3 consecutive readings from each observer) and single (from the first reading from each observer) measurement by the corneal dynamic Scheimpflug analyzer Corvis ST.

ParameterMean (μm) ± SD (μm)TRT (μm)CoV (%)ICC (95% CI)

Average536.6 ± 27.23.59.70.650.984
(0.973 to 0.990)
Single536.4 ± 27.54.512.60.850.973
(0.958 to 0.982)

SD = standard deviation, = within-subject standard deviation, TRT = test-retest repeatability (2.77), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.