Journal of Ophthalmology / 2015 / Article / Tab 4 / Research Article
Evaluation of Central Corneal Thickness Using Corneal Dynamic Scheimpflug Analyzer Corvis ST and Comparison with Pentacam Rotating Scheimpflug System and Ultrasound Pachymetry in Normal Eyes Table 4 Summary of previous studies for the agreement of CCT measurements obtained by other Scheimpflug systems, Orbscan II and specular microscopes in comparison to Pentacam or ultrasound pachymetry.
Author (year) Patients/eyes Device pairings Mean difference (μ m) ± SD value95% LoA (μ m) Lanza et al. [27 ] (2015) 102/102 Pentacam-Orbscan II 13.66 ± 16.53 <0.0001 −18.74 to 46.06 Sirius-Orbscan II 15.18 ± 17.16 <0.0001 −18.45 to 48.81 Sirius-Pentacam 1.52 ± 6.21 0.015 −10.65 to 13.69 Khaja et al. [28 ] (2015) 32/32 USP-Orbscan II 2.8 ± 0.28 NA −30.15 to 24.40 USP-specular microscopy 8.69 ± 1.24 NA −8.82 to 27.4 Smedowski et al. [29 ] (2014) 76/152 Corvis ST-Pentacam NA >0.05 NA Corvis ST-USP NA >0.05 NA Huang et al. [30 ] (2014) 66/66 Pentacam-Sirius −3.3 ± 5.2 <0.001 −13.6 to 6.9 Pentacam-Galilei −9.3 ± 3.7 <0.001 −16.6 to −2.0 Sirius-Galilei −6.0 ± 4.0 <0.001 −13.8 to 1.9 Anayol et al. [31 ] (2014) 32/32 Galilei-Pentacam 13.93 ± 0.88 <0.001 11.74 to 16.12 Galilei-Sirius 14.66 ± 0.69 <0.001 12.96 to 16.37 Pentacam-Sirius 0.73 ± 0.93 1.0 −1.50 to 3.02 Maresca et al. [32 ] (2014) 35/35 Sirius-USP −13.9 ± 14.4 <0.001 −42.2 to 14.4 Feizi et al. [33 ] (2014) 88/88 USP-Orbscan −14.5 ± 22.9 <0.001 −59.4 to 30.4 USP-Galilei −16.0 ± 19.6 <0.001 −54.5 to 22.5 Orbscan-Galilei −1.5 ± 17.0 0.99 −34.8 to 31.9
De La Parra-Colín et al. [34 ] (2014) 16/16 Sirius-Pentacam −10.1 ± 9.0 NA −27.7 to 7.5
Jorgel et al. [35 ] (2013) 50/50 Sirius-USP 4.68 ± 10.47 0.003 −15.84 to 25.20 Bayhan et al. [36 ] (2014) 50/50 USP-Sirius 17.58 ± 8.13 <0.001 15.27 to 19.89 Huang et al. [37 ] (2013) 43/43 Sirius-USP 6.88 ± 6.77 0.000 −6.39 to 20.14 Nassiri et al. [38 ] (2014) 32/61 Pentacam-USP −1 ± 9 0.32 −20 to 17
Orbscan II-USP 6 ± 14 <0.001 −21 to 33 Al Farhan et al. [39 ] (2013) 30/30 USP-specular microscopy −2.40 ± 9.10 0.16 −38.70 to 39.90 Tai et al. [40 ] (2013) 92/184 Pentacam-USP 10.08 ± 10.96 0.012 −11.40 to 31.56 Specular microscopy-USP −20.49 ± 8.91 <0.001 −37.95 to −3.04 Specular microscopy-Pentacam −30.57 ± 10.26 <0.001 −59.69 to −10.45 Chen et al. [41 ] (2012) 35/35 Pentacam-USP 5.27 ± 9.55 0.007 −24.0 to −13.4 Aramberri et al. [42 ] (2012) 35/35 Pentacam-Galilei −2.76 ± 4.52 <0.01 −6.1 to 11.6
González-Pérez et al. [43 ] (2011) 22/22 USP-Pentacam 3 ± 10 0.233 −16.2 to 21.2 USP-Orbscan II 32 ± 15 <0.001 3.1 to 60.8 USP-specular microscopy 26 ± 37 0.004 −46.2 to 97.8 Pentacam-Orbscan II 29 ± 11 <0.001 7.2 to 51.6 Pentacam-specular microscopy 23 ± 32 0.003 −40.2 to 86.2 Orbscan II-specular microscopy −6 ± 35 0.399 −74.8 to 62.0 USP-Orbscan II −15 ± 17 0.001 −47.9 to 18.7 Pentacam-Orbscan II −17 ± 14 <0.001 −43.9 to 9.8