Clinical Study

Repeatability and Reproducibility of Noninvasive Keratograph 5M Measurements in Patients with Dry Eye Disease

Table 2

Intraexaminer repeatability and interexaminer reproducibility of TMH and NIKBUTs in Healthy and DED groups.

ParametersHealthy ()Dry eye disease ()
Mean ± SDPrecisionRepeatabilityCV%ICC95% CI for ICC Mean ± SDPrecisionRepeatabilityCV%ICC95% CI for ICC

Intraexaminer repeatability
 TMH (mm)0.26 ± 0.120.100.1418.890.840.75 to 0.900.24 ± 0.080.080.1115.960.760.64 to 0.85
 NIKBUT first (s)7.25 ± 3.453.705.2426.100.750.62 to 0.845.77 ± 3.042.673.7623.540.820.73 to 0.89
 NIKBUT average (s)10.61 ± 3.933.965.6019.060.780.66 to 0.868.27 ± 3.863.044.2918.710.850.77 to 0.91
Interexaminer reproducibility
 TMH (mm)0.26 ± 0.150.100.1419.760.24 ± 0.080.080.1116.08
 NIKBUT first (s)7.35 ± 3.682.924.1320.305.87 ± 3.072.513.5521.85
 NIKBUT average (s)10.71 ± 4.113.655.1517.378.21 ± 3.932.783.9417.30

SD = standard deviation; TMH = tear meniscus height; NIKBUT = noninvasive Keratograph tear breakup time; CV = coefficient of variation; ICC = intraclass correlation coefficient; 95% CI = 95% confidence interval for the mean.