Clinical Study

Keraring Intrastromal Segment Depth Measured by Spectral-Domain Optical Coherence Tomography in Eyes with Keratoconus

Table 2

Keraring measured and intended distance from the anterior and the posterior corneal surfaces.

CaseEyeIntended distance
(μm)
Distance apex/
anterior surface (μm)
Distance inner corner/
posterior surface (μm)
Distance outer corner/
posterior surface (μm)
Anterior surface Posterior surfaceMeasuredAbsolute difference MeasuredAbsolute differenceMeasuredAbsolute difference

1RE391136205186983822589
2LE333176287461205620731
335199313221465321011
3LE3501622589215210267105
4LE3701473066411928328181
5RE39912220119815432260138
6LE37712521616115732266141
7RE40518025015516911285105
8RE4001003029898219898
LE3581183431510117239121
9RE36614217119518139318176
10RE34411816717714325285167
LE35712121514215433260139
11RE399145328711301521772
12RE40015230311717321314162
LE40116019820317616359199
13RE323164270531194519733
14LE3561252551011261251126
LE35419032133190024555
15LE362902966687316676
16LE360156349111451117216
17RE39215918820417516329170
18LE38312031865982214222
19RE3589027385791115868
20RE38712025113615636289169
21RE40012526413616439272147

MEAN371.54140.08263.38108.92138.8523.54248.42108.35
SD24.4228.9354.1662.7231.7815.9057.3256.99
RANGE323–40590–199167–34911–20479–1900–56142–35911–199

RE: right eye. LE: left eye. Eyes with intrastromal tunnel created using femtosecond laser. SD: Standard Deviation.