Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices

Table 1

CCT and keratometry measurements from all 4 devices.

ParameterDevice
IOLMaster 700Pentacam HRiDesignCirrus HD-OCT

CCT [μm]
 Mean ± SD543.20 ± 31.69554.19 ± 26.86542.75 ± 31.34
 Range472–616484–614468–628
J0 [D]
 Mean ± SD0.350 ± 0.3670.384 ± 0.3420.358 ± 0.353
 Range−0.65–1.32−0.51–1.18−0.58–1.21
J45 [D]
 Mean ± SD−0.027 ± 0.197−0.385 ± 0.218−0.023 ± 0.205
 Range−0.55–0.44−0.73–0.41−0.56–0.47
Kmean [D]
 Mean ± SD42.76 ± 1.4943.25 ± 1.5143.52 ± 1.54
 Range38.89–45.9039.42–46.4339.54–46.65

CCT: central corneal thickness; J0: astigmatism 90/180 degrees; J45: astigmatism 45/135 degrees; Kmean: mean keratometry; —: no data available because of device specifications.