Research Article

Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices

Table 2

Intraoperator repeatability measurements.

Parameter/deviceICC95% confidence intervalDifferencea (mean ± SD)95% LoA value

CCT [μm]
 IOLMaster 7000.9990.998–0.999−0.09 ± 2.76−0.83–0.650.808
 Pentacam HR0.9950.992–0.997−0.25 ± 5.18−1.65–1.140.717
 Cirrus HD-OCT0.9950.992–0.9970.91 ± 5.79−0.65–2.470.250
J0 [D]
 IOLMaster 7000.9890.983–0.9930.02 ± 0.11−0.01–0.040.213
 Pentacam HR0.9860.978–0.991−0.02 ± −0.12−0.05–0.010.252
 iDesign0.9940.991–0.996−0.007 ± 0.07−0.007–0.010.468
J45 [D]
 IOLMaster 7000.9740.959–0.9840.009 ± 0.08−0.01–0.030.406
 Pentacam HR0.9840.975–0.990−0.02 ± 0.06−0.03–0.0010.066
 iDesign0.9770.963–0.986−0.01 ± 0.08−0.03–0.010.304
Kmean [D]
 IOLMaster 7000.9990.999–1.0−0.002 ± 0.09−0.03–0.020.842
 Pentacam HR0.9990.999–1.0−0.02 ± 0.08−0.03–0.0050.145
 iDesign0.9920.987–0.9950.05 ± 0.29−0.03–0.130.211

CCT: central corneal thickness; J0: astigmatism 90/180 degrees; J45: astigmatism 45/135 degrees; Kmean: mean keratometry; ICC: intraclass correlation coefficient; LoA: limits of agreement; abetween the second measurement of each eye; values were calculated by one-sample t-test.