Clinical Study

Visual Outcomes, Quality of Vision, and Quality of Life of Diffractive Multifocal Intraocular Lens Implantation after Myopic Laser In Situ Keratomileusis: A Prospective, Observational Case Series

Table 2

Intraocular lens formula accuracy (25 eyes with pre-laser in situ keratomileusis data available).

FormulaPredicted error in refraction (dioptre)
NumericalAbsolute
Mean ± SDRangeMean ± SDMedianRange

Clinical history method
 With SRK/T−2.72, 1.250.450.02, 2.72
 With Haigis−5.17, 0.261.080.02, 5.17
Using Refractive Change
 Modified-Masket−2.55, 0.450.410.03, 2.55
 Barrett True K−2.63, 0.610.340.02, 2.63
Using no prior data
 Shammas−3.78, 0.630.440.05, 3.78
 Haigis-L−3.16, 0.090.320.01, 3.16
 Barrett True K No History−3.45, 0.750.260.01, 3.45

Comparison between clinical history method with Haigis formula and the remaining formulas (clinical history method with SRK/T, ; Modified-Masket formula, ; Barrett True K, ; Haigis-L, ; Shammas, ; and Barrett True K, ), where was considered statistically significant for multiple comparisons. Comparison between the remaining formulas ( for all comparisons).
SD = standard deviation.