Research Article
Panretinal Photocoagulation Using Short-Pulse Laser Induces Less Inflammation and Macular Thickening in Patients with Diabetic Retinopathy
Table 3
Summary of parameters relating to lasers.
| | Conv group | SPL group |
| Power (mW) | 257 ± 83 | 413 ± 107 | Pulse duration (ms) | 200 | 20 | Spot size (um) | 200 | 200 | Wavelength (nm) | 532 | 532 | Total energy/spot (mJ) | 51.4 ± 16.6 | 8.3 ± 10.4 | Number of spots | 1203 ± 59 | 2925 ± 125 | Type of laser spots | Single spot | Pattern spot (5 × 5, 4 × 4 arrays) | Complications | | | Vitreous hemorrhage (%) | 1 (4.8) | 1 (4.8) | Neovascular glaucoma (%) | 0 (0) | 0 (0) |
|
|
Conv: conventional; SPL: short-pulse laser.
|