Research Article

Applying a New Automated Perimetry Pattern Based on the Stimulus Distribution of the Multifocal ERG to Improve Structure-Function Investigation in Glaucoma

Figure 1

Superposition of the mf103 pattern and the mfERG stimulus grid (103 hexagons). The protocol has been developed to stimulate the individual areas, covering 50°, using the same test parameters as in standard perimetry testing: size III/3e; 100 ms duration; background 10 cd/m2; 0 dB equaled to 4000 asb.