Research Article

Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses

Table 3

Comparison among normal subjects and slight and other subluxations of 25 MHz UBM.

Distance (mm)Normal subjects ()Slight subluxation ()Other subluxations ()

Maximum1.01 ± 0.201.12 ± 0.292.06 ± 0.810.000
Minimum0.38 ± 0.290.41 ± 0.490.800
D-value0.01 ± 0.010.76 ± 0.271.58 ± 0.760.000

All measuring values are presented as means ± SD along with 95% confidence intervals (CIs). Maximum = the maximum distance between lens equator and ciliary process of the subluxated lens; minimum = the minimum distance between lens equator and ciliary process of the subluxated lens; D-value = the difference between maximum and minimum distance of the subluxated lens. .