Research Article
Comparison of Color Light-Emitting Diode Corneal Topographer and Dual Rotating Scheimpflug–Placido Topographer
Table 4
Comparison of corneal index measurements using dual rotating Scheimpflug–Placido and color light-emitting diode corneal topographers in total group.
| Parameter | Total group | DRSP (mean ± SD) | Color LED (mean ± SD) | Difference | value | ICC | Mean ± SD | 95% CI |
| Anterior corneal indices (D) | simK | 41.07 ± 2.76 | 41.09 ± 2.87 | 0.03 ± 0.37 | -0.04, 0.09 | 0.451 | 0.996 | Astigmatism magnitude | 1.09 ± 0.72 | 1.13 ± 0.78 | 0.04 ± 0.36 | −0.03, 0.10 | 0.256 | 0.938 | J0 | −0.46 ± 0.39 | −0.47 ± 0.44 | −0.00 ± 0.19 | −0.04, 0.03 | 0.874 | 0.947 | J45 | 0.04 ± 0.24 | −0.02 ± 0.25 | −0.05 ± 0.15 | −0.08, 0.03 | 0.000 | 0.906 | Posterior corneal indices (D) | simK | −6.41 ± 0.22 | −6.12 ± 0.32 | 0.29 ± 0.27 | 0.24, 0.34 | 0.000 | 0.681 | Astigmatism magnitude | −0.36 ± 0.11 | −0.40 ± 0.19 | −0.04 ± 0.20 | −0.07, −0.00 | 0.027 | 0.334 | J0 | 0.17 ± 0.06 | 0.13 ± 0.13 | −0.04 ± 0.14 | −0.06, −0.01 | 0.004 | 0.180 | J45 | −0.01 ± 0.06 | −0.01 ± 0.12 | −0.00 ± 0.13 | −0.03, 0.02 | 0.695 | 0.000 |
|
|
CI, confidence interval; color LED, color light-emitting diode corneal topographer; DRSP, dual rotating Scheimpflug–Placido corneal topographer; ICC, intraclass correlation coefficient; K, keratometry. Paired sample t-test. |