Clinical Study
Prospective Randomized Multicenter Comparison of the Clinical Outcomes of V4c and V5 Implantable Collamer Lenses: A Contralateral Eye Study
Table 2
Comparison of visual and refractive outcomes between the V4c and V5 models 6 months after surgery.
| Variable | V4c | V5 | value |
| UCDVA (logMAR) (Snellen) | −0.23 ± 0.09 20/12 | −0.19 ± 0.12 20/13 | 0.152 | CDVA(logMAR) (Snellen) | −0.25 ± 0.07 20/11 | −0.26 ± 0.07 20/11 | 0.500 | Manifest sphere (D) | 0.20 ± 0.30 | −0.22 ± 0.42 | 0.497 | Manifest cylinder (D) | −0.29 ± 0.29 | −0.27 ± 0.22 | 0.125 | Spherical equivalent (D) | 0.05 ± 0.07 | −0.09 ± 0.38 | 0.505 | Efficacy index | 1.16 ± 0.22 | 1.03 ± 0.23 | 0.044 | Safety index | 1.21 ± 0.20 | 1.19 ± 0.20 | >0.999 |
|
|
UCDVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity. Values are presented as mean ± standard deviation.
|