Clinical Study

Prospective Randomized Multicenter Comparison of the Clinical Outcomes of V4c and V5 Implantable Collamer Lenses: A Contralateral Eye Study

Table 2

Comparison of visual and refractive outcomes between the V4c and V5 models 6 months after surgery.

VariableV4cV5 value

UCDVA (logMAR)
(Snellen)
−0.23 ± 0.09
20/12
−0.19 ± 0.12
20/13
0.152
CDVA(logMAR)
(Snellen)
−0.25 ± 0.07
20/11
−0.26 ± 0.07
20/11
0.500
Manifest sphere (D)0.20 ± 0.30−0.22 ± 0.420.497
Manifest cylinder (D)−0.29 ± 0.29−0.27 ± 0.220.125
Spherical equivalent (D)0.05 ± 0.07−0.09 ± 0.380.505
Efficacy index1.16 ± 0.221.03 ± 0.230.044
Safety index1.21 ± 0.201.19 ± 0.20>0.999

UCDVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity. Values are presented as mean ± standard deviation.