Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST

Table 1

Descriptive data of participants and results of intraocular pressure (IOP) measurements (mean value ± standard deviation).

Descriptive data

Number of subjects94
Eye (right/left)47 (50%)/47 (50%)
Gender (female/male)39 (41.5%)/55 (58.5%)
Mean age (years)45.6 ± 17.2
CCT (μm)553 ± 33
Intraocular pressure
GAT-IOPc (mmHg)12.9 ± 2.4
DCT-IOP (mmHg)16.1 ± 2.6
IOPcc (mmHg)15.6 ± 3.3
bIOP (mmHg)13.5 ± 2.4

GAT-IOPc: corrected Goldmann applanation tonometry-IOP; DCT-IOP: dynamic contour tonometry-IOP; IOPcc: corneal compensated IOP by using ocular response analyzer; bIOP: biomechanical corrected IOP by using Corvis ST; CCT: central corneal thickness.