OCT Biometry (B-OCT): A New Method for Measuring Ocular Axial Dimensions

Table 1

Intraobserver repeatability results for B-OCT using REVO NX based on three measurements taken by 2 clinicians in a sample of 50 eyes.

ParameterObserverMean ± SDSwTRTCoV (%)ICC

AXL (mm)1st23.93 ± 0.800.010.030.041.000
2nd23.93 ± 0.800.010.030.041.000

CCT (μm)1st555.44 ± 26.562.015.580.360.994
2nd555.43 ± 27.081.754.850.320.996

ACD (mm)1st3.59 ± 0.280.010.040.380.998
2nd3.59 ± 0.280.010.030.340.998

LT (mm)1st4.02 ± 0.220.020.050.450.993
2nd4.02 ± 0.220.010.040.350.996

AXL: axial length; CCT: central corneal thickness; ACD: anterior chamber depth; LT: lens thickness; Sw: within-subject SD; TRT, test-retest repeatability (2.77 Sw); CoV: within-subject coefficient of variation; ICC: intraclass correlation coefficient.