OCT Biometry (B-OCT): A New Method for Measuring Ocular Axial Dimensions

Table 3

Measurement values comparison between B-OCT REVO NX and IOLMaster 700.

nDeviceMeanSDPaired samples t-test ( value)RangeDifference of the meansSD for difference of the means95% CI for difference of the meansICC

Healthy
 Axial length (mm)164REVO NX23.561.560.266419.11–34.48−0.0010.016−0.004 to 0.0011.000
IOLMaster 70023.561.5619.09–34.48
 Anterior chamber depth (mm)164REVO NX3.350.340.92651.65–4.320.0000.024−0.004 to 0.0040.998
IOLMaster 7003.350.351.61–4.29
 Lens thickness (mm)164REVO NX4.080.410.33983.30–5.34−0.0020.024−0.006 to 0.0020.998
IOLMaster 7004.080.413.31–5.37
 Central corneal thickness (µm)164REVO NX554.134.40.0552478.1–682.3−0.85.1−1.6 to 0.00.989
IOLMaster 700553.435.5477.3–684.0

Cataract
 Axial length (mm)111REVO NX23.811.890.118519.62–32.07−0.0030.022−0.007 to 0.0011.000
IOLMaster 70023.801.8919.60–32.01
 Anterior chamber depth (mm)112REVO NX3.100.410.33262.16–3.980.0030.029−0.003 to 0.0080.997
IOLMaster 7003.100.412.14–4.04
 Lens thickness (mm)111REVO NX4.640.350.43643.94–5.48−0.0020.025−0.007 to 0.0030.997
IOLMaster 7004.630.353.94–5.51
 Central corneal thickness (µm)115REVO NX546.533.1<0.0001437.8–631.02.76.41.5 to 3.90.979
IOLMaster 700549.233.6438.2–647.0

Difference of the means was computed by subtracting REVO NX values from IOLMaster 700 values.