Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System

Table 3

Comparison of corneal astigmatism values between devices.

DeviceCorrelation between devices ()Mean difference (D) ± SD95% CI95% of LoA spread (D)
MagnitudeCardinal componentOblique componentMagnitudeCardinal componentOblique componentMagnitudeCardinal componentOblique componentMagnitudeCardinal componentOblique component

IOLMaster vs. iTrace0.853 (<0.0001)0.965 (<0.0001)0.920 (<0.0001)0.03 ± 0.47−0.09 ± 0.480.08 ± 0.34−0.07, 0.13−0.19, 0.010.01, 0.151.841.891.32
IOLMaster vs. Pentacam0.870 (<0.0001)0.957 (<0.0001)0.839 (<0.0001)0.05 ± 0.450.07 ± 0.530.00 ± 0.50−0.04, 0.15−0.04, 0.18−0.10, 0.111.772.081.97
iTrace vs. Pentacam0.881 (<0.0001)0.967 (<0.0001)0.868 (<0.0001)0.02 ± 0.420.16 ± 0.46−0.08 ± 0.45−0.07, 0.110.07, 0.26−0.17, 0.011.661.781.76

CI = confidence interval; D = diopters; LoA = limit of agreement; SD = standard deviation.