Research Article

Repeatability of Aberrometry-Based Automated Subjective Refraction in Healthy and Keratoconus Subjects

Table 4

Mean difference between sessions (bias) and its standard deviation (SD) for spherical equivalent (M), cylindrical vectors (J0 and J45), binocular corrected visual acuity (BCDVA), and spent time in refraction.

VariableGroupSession 1-session 2Session 1–session 3Session 2-session 3ANOVA ( value)

M (D)HealthyBias ± SD0.00 ± 0.07−0.04 ± 0.08−0.04 ± 0.050.806
value>0.999>0.999>0.999
KeratoconusBias ± SD−0.32 ± 0.18−0.35 ± 0.17−0.03 ± 0.130.146
value0.2450.166>0.999

J0 (D)HealthyBias ± SD−0.01 ± 0.02−0.02 ± 0.02−0.01 ± 0.030.607
value>0.999>0.999>0.999
KeratoconusBias ± SD0.07 ± 0.070.08 ± 0.080.01 ± 0.080.511
value0.9130.944>0.999

J45 (D)HealthyBias ± SD0.00 ± 0.020.01 ± 0.020.01 ± 0.020.756
value>0.999>0.999>0.999
KeratoconusBias ± SD−0.06 ± 0.05−0.05 ± 0.070.02 ± 0.060.518
value0.748>0.999>0.999

BCDVA (logMAR)HealthyBias ± SD0.01 ± 0.010.02 ± 0.010.00 ± 0.010.425
P value>0.9990.621>0.999
KeratoconusBias ± SD0.01 ± 0.01−0.02 ± 0.01−0.02 ± 0.010.143
value>0.9990.7070.142

Time (min : s)HealthyBias ± SD0 : 22 ± 0.070 : 23 ± 0 : 090 : 01 ± 0 : 050.011
value0.007¥0.037¥>0.999
KeratoconusBias ± SD0 : 20 ± 0 : 080 : 30 ± 0 : 130 : 10 ± 0 : 110.047
value0.0800.081>0.999

The statistical comparison was performed between sessions. , one-way ANOVA for related samples. ¥, pairwise comparison with Bonferroni correction.