Research Article
First Experience in Small Incision Lenticule Extraction with the Femto LDV Z8 and Lenticule Evaluation Using Scanning Electron Microscopy
Table 1
Main equipment parameters of the two femtosecond laser platforms.
| | Femto LDV Z8 | VisuMax |
| Manufacturer | Ziemer Ophthalmic Systems AG, Port, Switzerland | Carl Zeiss Meditec, Jena, Germany | Wavelength (nm) | 1020–1060 | 1043 | Pulse duration (fs) | 200–350 | 220–580 | Laser pulse repetition rate (kHz) | >5000 | 500 | Weight (kg) | 215 | 870 | Docking method | Floating a mobile hand-piece on eye | Sliding patient bed under a fixed laser head | Contact glass on suction system | Flat | Curve | Interoperative OCT | Yes | No | Automatic detection of pupil | Yes | No | Pupil central offsetting | Yes | No | Real-time video recording | No | Yes |
|
|