Comparison of Two Wavefront Autorefractors: Binocular Open-Field versus Monocular Closed-Field

Table 3

Intersession repeatability of mean spherical equivalent (MSE) obtained with both binocular open-field (eye refract) and monocular closed-field (VX110) autorefractors in terms of mean difference between sessions (bias), its standard deviation (SD), repeatability (Sr), and its 95% limit (r).

Wavefront autorefractorGroupMSESession 1-session 2Session 1–session 3Session 2-session 3Repeatability [95% limit] (D)

Eye refractTotalBias ± SD (D)−0.02 ± 0.330.00 ± 0.280.02 ± 0.290.21 [0.59]
value1.0001.0001.000
TeenagersBias ± SD (D)−0.08 ± 0.16−0.10 ± 0.24−0.02 ± 0.190.14 [0.40]
value0.1020.1881.000
AdultsBias ± SD (D)−0.02 ± 0.21−0.02 ± 0.230.00 ± 0.200.15 [0.41]
value1.0001.0001.000
PresbyopesBias ± SD (D)0.01 ± 0.450.06 ± 0.320.04 ± 0.370.27 [0.75]
value1.0000.6921.000

VX110TotalBias ± SD (D)−0.07 ± 0.40−0.05 ± 0.870.02 ± 0.880.53 [1.47]
value0.3251.0001.000
TeenagersBias ± SD (D)−0.18 ± 0.770.18 ± 1.520.36 ± 1.490.92 [2.55]
value0.9061.0000.843
AdultsBias ± SD (D)−0.06 ± 0.28−0.09 ± 0.22−0.04 ± 0.230.18 [0.50]
value0.7600.0711.000
PresbyopesBias ± SD (D)−0.02 ± 0.17−0.12 ± 0.77−0.10 ± 0.760.44 [1.23]
value1.0000.9331.000

The statistical comparison was done between sessions. , one-way ANOVA for paired samples with Bonferroni correction.