Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients

Table 1

Intraobserver repeatability outcomes for biometric measurements obtained using IOLMaster 700 swept-source optical coherence tomography in children.

ParameterObserverMean ± SDSwTRTCoV (%)ICC (95% CI)

AL (mm)1st24.7 ± 1.070.010.030.041.000 (1.000 to 1.000)
2nd24.7 ± 1.070.010.020.031.000 (1.000 to 1.000)
CCT (μm)1st552.27 ± 33.62.807.760.510.993 (0.990 to 0.995)
2nd551.78 ± 33.22.867.930.520.993 (0.989 to 0.995)
ACD (mm)1st3.76 ± 0.240.020.050.480.994 (0.992 to 0.996)
2nd3.76 ± 0.230.010.040.400.996 (0.994 to 0.997)
LT (mm)1st3.38 ± 0.150.020.070.720.973 (0.962 to 0.981)
2nd3.38 ± 0.150.020.060.610.981 (0.974 to 0.987)
Km (D)1st42.73 ± 1.550.090.260.220.996 (0.995 to 0.997)
2nd42.73 ± 1.550.110.300.250.995 (0.993 to 0.997)
WTW (mm)1st12.28 ± 0.430.090.260.750.955 (0.938 to 0.968)
2nd12.27 ± 0.420.090.260.760.952 (0.934 to 0.966)

AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, LT = lens thickness, K = keratometry, WTW = white to white, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, and ICC = intraclass correlation coefficient.