Research Article

Corneal and Epithelial Thickness Mapping: Comparison of Enhanced Spectral-Domain- and Spectral-Domain-Optical Coherence Tomography

Table 1

Interobserver reproducibility of corneal epithelial thickness assessment with Enhanced Spectral-Domain and Spectral-Domain-Optical Coherence Tomography in normal eyes.

Mean standard deviation (μm)Mean coefficient of variation (%)
ESD-OCTSD-OCTESD-OCTSD-OCT

Central0.880.730.341.71.30.24
Minimal1.181.170.982.42.40.97
Maximal1.061.0320.891.81.80.96
2–5 mm superior1.790.870.0013.41.60.0006
2–5 mm supero-nasal1.590.760.00043.01.40.0002
2–5 mm nasal1.460.820.0022.71.50.001
2–5 mm infero-nasal1.140.750.022.11.40.01
2–5 mm inferior1.360.770.0022.51.40.001
2–5 mm infero-temporal1.360.690.0022.51.30.001
2–5 mm temporal1.440.700.00012.81.30.00005
2–5 mm supero-temporal1.670.930.0053.21.80.003
5–6 mm superior2.821.060.0025.72.10.002
5–6 mm supero-nasal2.160.930.0000074.11.80.000004
5–6 mm nasal1.430.870.0012.61.60.002
5–6 mm infero-nasal1.660.860.0033.01.60.004
5–6 mm inferior1.460.750.0022.61.40.002
5–6 mm infero-temporal1.560.830.0032.81.60.006
5–6 mm temporal1.880.910.00063.61.70.0009
5–6 mm supero-temporal1.791.220.0453.52.40.045

Each measurement was performed by 3 trained operators in 30 eyes of 30 patients. Data in bold indicate significant differences between ESD-OCT and SD-OCT.