Research Article

Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions

Table 2

Occurrence of IOL delivery system interactions for corneal incisions.

Delivery system and knife size (mm)IOL adherence to plungerSudden ejection of IOLOverriding plunger within cartridgeIntrawound lens manipulationTrapped trailing hapticIOL behaviorGross damage to IOLIncision enlargement ≥0.2 mmCartridge tip splitting

A 2.20/100/101/100/101/100/100/105/100/10
A 2.40/100/101/100/100/100/100/100/100/10
iT 2.20/100/100/103/100/100/100/109/100/10
iT 2.41/100/100/103/100/100/100/102/100/10
iS 2.00/100/100/100/100/100/100/1010/1010/10

IOL = intraocular lens; A = AutonoMe; iT = iTec; iS = iSert.