Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions
Table 3
Occurrence of IOL delivery system interactions for sclerocorneal incisions.
| Delivery system and knife size (mm) | IOL adherence to plunger | Sudden ejection of IOL | Overriding plunger within cartridge | Intrawound lens manipulation | Trapped trailing haptic | IOL behavior | Gross damage to IOL | Incision enlargement ≥0.2 mm | Cartridge tip splitting |
| A 2.2 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 3/10 | 0/10 | A 2.4 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 1/10 | 0/10 | iT 2.2 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 8/10 | 0/10 | iT 2.4 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 8/10 | 0/10 | iS 1.8 | 0/10 | 0/10 | 0/10 | 1/10 | 0/10 | 0/10 | 0/10 | 10/10 | 10/10 |
|
|
IOL = intraocular lens; A = AutonoMe; iT = iTec; iS = iSert.
|