Research Article

Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions

Table 4

Mean enlargement and final size of corneal incisions.

Device (mm)A 2.2 mmA 2.4 mmiT 2.2 mmiT 2.4 mmiS 2.0 mm

Mean incision enlargement0.14 ± 0.070.07 ± 0.050.22 ± 0.060.10 ± 0.070.24 ± 0.05
Mean final incision size2.36 ± 0.072.48 ± 0.062.48 ± 0.082.60 ± 0.072.29 ± 0.09

A = AutonoMe; iT = iTec; iS = iSert. Data are represented as means ± SD.