Research Article

Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions

Table 5

Mean enlargement and final size of sclerocorneal incisions.

DeviceA 2.2 mmA 2.4 mmiT 2.2 mmiT 2.4 mmiS 1.8 mm

Mean incision enlargement (mm)0.13 ± 0.050.08 ± 0.060.21 ± 0.070.17 ± 0.070.35 ± 0.08
Mean final incision size (mm)2.32 ± 0.062.49 ± 0.032.59 ± 0.092.66 ± 0.082.34 ± 0.11

A = AutonoMe; iT = iTec; iS = iSert. Data are represented as means ± SD.