One-Year Visual and Refractive Outcomes following LASIK for Myopia and Myopic Astigmatism with MEL 90 versus Schwind Amaris 750S Excimer Laser: A Comparative Study

Table 2

The postoperative visual and refractive results obtained at 1 year for both study groups.

Parameter (mean ± SD) (range)Schwind Amaris 750SMEL 90 value

Sphere (D)−0.15 ± 0.20 (−0.75 to 1.25)−0.15 ± 0.21 (−1.00 to 0.50)1.00
Cylinder (D)−0.13 ± 0.18 (−0.75 to 0.5)−0.14 ± 0.21 (−1.00 to 0.00)0.79
SE (D)−0.21 ± 0.22 (−0.87 to 1.25)−0.23 ± 0.23 (−1.00 to 0.25)0.29
UDVA (logMAR)−0.10 ± 0.09 (−0.2 to 0.2)−0.05 ± 0.07 (−0.20 to 0.10)0.24
CDVA (logMAR)−0.14 ± 0.06 (−0.2 to 0)−0.09 ± 0.08 (−0.20 to 0.00)0.24
Safety index1.12 ± 0.16 (0.62 to 1.6)1.08 ± 0.15 (0.78 to 1.6)0.29
Efficacy index1.02 ± 0.10 (0.63 to 1.28)1.00 ± 0.10 (0.5 to 1.25)0.90
AOE (arithmetic)0.29 ± 6.05 (−24 to 30)0.37 ± 7.34 (−30 to 40)0.96
AOE (absolute)2.87 ± 5.33 (0 to 30)3.22 ± 6.60 (0 to 40)0.82
CI0.95 ± 0.33 (0 to 2.02)0.93 ± 0.26 (0.21 to 1.48)0.80

SD: standard deviation; SE: spherical equivalent; UDVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity; AOE: angle of error; CI: correction index.