Journal of Ophthalmology / 2021 / Article / Tab 2 / Research Article
One-Year Visual and Refractive Outcomes following LASIK for Myopia and Myopic Astigmatism with MEL 90 versus Schwind Amaris 750S Excimer Laser: A Comparative Study Table 2 The postoperative visual and refractive results obtained at 1 year for both study groups.
Parameter (mean ± SD) (range) Schwind Amaris 750S MEL 90 valueSphere (D) −0.15 ± 0.20 (−0.75 to 1.25) −0.15 ± 0.21 (−1.00 to 0.50) 1.00 Cylinder (D) −0.13 ± 0.18 (−0.75 to 0.5) −0.14 ± 0.21 (−1.00 to 0.00) 0.79 SE (D) −0.21 ± 0.22 (−0.87 to 1.25) −0.23 ± 0.23 (−1.00 to 0.25) 0.29 UDVA (logMAR) −0.10 ± 0.09 (−0.2 to 0.2) −0.05 ± 0.07 (−0.20 to 0.10) 0.24 CDVA (logMAR) −0.14 ± 0.06 (−0.2 to 0) −0.09 ± 0.08 (−0.20 to 0.00) 0.24 Safety index 1.12 ± 0.16 (0.62 to 1.6) 1.08 ± 0.15 (0.78 to 1.6) 0.29 Efficacy index 1.02 ± 0.10 (0.63 to 1.28) 1.00 ± 0.10 (0.5 to 1.25) 0.90 AOE (arithmetic) 0.29 ± 6.05 (−24 to 30) 0.37 ± 7.34 (−30 to 40) 0.96 AOE (absolute) 2.87 ± 5.33 (0 to 30) 3.22 ± 6.60 (0 to 40) 0.82 CI 0.95 ± 0.33 (0 to 2.02) 0.93 ± 0.26 (0.21 to 1.48) 0.80
SD: standard deviation; SE: spherical equivalent; UDVA: uncorrected distance visual acuity; CDVA: corrected distance visual acuity; AOE: angle of error; CI: correction index.