Research Article
Sensitivity Improvement of an Impedimetric Immunosensor Using Functionalized Iron Oxide Nanoparticles
Table 5
Comparison of analytical parameters of immunosensors based on SAM and immunosensor based on iron oxide nanoparticles.
| | Analytical parameters | SAM system | IrNp System |
| Optical characterization | detection limit | 4.5 g/ml | 300 ng/ml | sensitivity | | | Dynamic range | 4.5 to 15 g/ml | 0.3 to 1 g/ml | The time response | 45 minutes | 30 minutes | reproducibility | 6.3% | 5.7% |
| Electrical characterization | detection limit | 1 ng | 500 pg | sensitivity | | | Dynamic range | 1 to 50 ng | 0.5 to 50 ng | The time response | 45 minutes | 20 minutes | reproducibility | 8.7% | 7.5% |
|
|