Abstract

Optical fiber sensors using luminescent probes located along an optical fiber in the cladding of this fiber are of great interest for monitoring physical and chemical properties in their environment. The interrogation of a luminophore with a short laser pulse propagating through the fiber core allows for the measurement of the location of these luminophores. To increase the spatial resolution of such a measurements and to measure multiple analytes and properties in a confined space, a crossed optical fiber sensing platform can be employed. Here we describe the application of this platform to measuring the concentration of dissolved oxygen. The sensor is based on luminescence quenching of a ruthenium complex immobilized in a highly crosslinked film and covalently attached to the optical fibers. Both luminescence-intensity and luminescence-lifetime changes of the sensor molecules in response to changes in the concentration of oxygen dissolved in water are reported. For luminescence-intensity measurements, a second adjacent sensor region is employed as reference to account for laser pulse energy fluctuations. Enhanced quenching response in water is demonstrated by the use of organically modified poly(ethylene glycol) precursors, which increase the hydrophobicity of the film surface.

1. Introduction

Optical fiber sensors using luminescent probes located along an optical fiber in the fiber cladding and in close proximity of the fiber core have been of great interest for spatially resolved measurement of physical and chemical properties along a single fiber in the sensor environment [13]. The location of an emitting luminophore along a fiber can be determined by measuring the time delay between a short exciting laser pulse propagating in the fiber core to the sensor site and the subsequent sensor luminescence pulse returning to the front end of this fiber. The spatial resolution of these measurements is limited by the sensor luminescence lifetimes, because a minimum separation of the luminophores is required to resolve returning light pulses. However, by using a second fiber, which forms orthogonal junctions with the first fiber at all sensor sites, as an optical delay line, the minimum spacing between adjacent sensor regions can be well below the luminescence lifetime limit [4]. Because the coupling between the two fibers is evanescent, the attenuation of the excitation pulse is low, making long arrays of sensor regions feasible.

To test the performance of the crossed-fiber sensor arrays for sensing in aqueous environments, we fabricated a crossed-fiber dissolved-oxygen sensor. Dissolved oxygen (DO) is an important parameter in biological, biomedical, and industrial systems. This widespread need has led to the development of a number of water-soluble, noninvasive, sensitive, and selective detection methods that are capable of real-time DO sensing. In the past, oxygen sensors have mainly been based on amperometric Clark-Type electrodes [5]. Within the past few years, research in optical sensors have overcome most of the disadvantages of amperometric sensors; in contrast to the latter, optical sensors are free of electromagnetic interference, do not consume oxygen and, in general, respond faster [69].

Different optical sensors have been reported [1016] for oxygen measurement in both gas and aqueous environments. Most of these sensors respond to changes in the luminescence intensity of an indicator dye; however, as with all measurements based on luminescence-intensity changes, these measurements are hampered by the fact that signal changes may arise from a number of factors other than the measurand. This problem could be overcome either by using luminescence-lifetime detection (using either direct time-resolved detection or the phase-shift method) [12, 1719] or by employing ratiometric sensors [20, 21]. In the latter case the ratio of the indicator-luminescence intensity and a reference-luminescence intensity can compensate for system fluctuations; in spite of the need for a reference, measurements based on indicator-luminescence changes are considered preferable to measurements based on luminescence-lifetime changes, because the latter require a more sophisticated experimental setup.

The use of luminescent ratiometric sensors combined with the optical fiber technology provides an opportunity of real-time and remote DO measurement. Fiber-optic oxygen sensors are frequently of the extrinsic type [2228], with a single sensor mounted at the distal end of the fiber. This allows for point monitoring; however, for measurements at multiple locations, separate fibers are needed for each additional sensor. Thus, placing luminescent sensors placed into the fiber cladding (or a suitable material replacing the original fiber cladding) along a continuous stretch of fiber (distributed sensing) or in many discrete regions (quasidistributed sensing) constitutes an efficient alternative; when appropriate fiber-guiding conditions are maintained at the sensor regions [4], the interaction of the probe light propagating in the fiber core and the luminosensor in the cladding is solely through the evanescent fields of the core modes.

The DO sensor implemented here consists of the sensor dye tris(1,10-phenanthroline)ruthenium(II) chloride, Ru[(Phen)3]Cl2, immobilized in a poly(ethylene glycol) (PEG) matrix. This polymer replaces the optical cladding, which is removed prior to covalent attachment of the PEG-film. The immobilization of this ruthenium complex within an optically transparent, low-refractive index and oxygen-permeable structural matrix is of primary concern for luminescent oxygen fiber-based sensors. Issues regarding oxygen permeability and luminophore solubility within the matrix, luminophore leaching, and complex-dependent luminescent quenching were considered. We measured sensor calibration curves and characterized the sensor with respect to response time, detection limit, repeatability, biocompatibility, longevity, and possible interferences.

2. Background

Our optical sensor consists of an optical fiber with exposed core, PEG-films containing the sensor and reference molecules, and a sensing medium. The versatility of the PEG-polymer synthesis enables optimization of sensor parameters (e.g., sensitivity, response time, and signal-to-noise ratio) via control of film properties such as coating length, thickness, and porosity. The PEG films replace the original cladding material only at the sensor/reference regions. The refractive index of the cladding is such that the guiding condition of the fiber is maintained, that is, total internal reflection of the light propagating in the fiber core is taking place at the core/cladding interface at the sensor regions. The interaction of light propagating in the fiber core, which is used to interrogate the sensor, and the indicator molecules in the cladding are through evanescent fields, which decay exponentially with distance from the core/cladding interface. This defines a short-range sensing volume within which the evanescent field may interact with the indicator molecules [29]. Sensor molecules within the range of evanescent fields may absorb light of the appropriate frequency, attenuating the total internal reflection [29]. Also, the luminescence of an indicator molecule may be coupled back into a guiding mode of the fiber via the evanescent fields.

The spatial information regarding the location of an emitting luminophore along the fiber can be obtained using optical time-domain reflectometry (OTDR) [3033] or, more appropriately for luminescence transduction, optical time-of-flight detection (OTOF) [34]. Laser pulses fed into the front end of the fiber cause pulsed evanescent excitation of the luminophores [35]. The luminescence pulses returning to the front of the fiber arrive with a time delay Δ𝑡𝑑 that allows for the calculation of the location 𝐿 of the emitting luminophore. In our setup, the luminescence is captured by a second fiber (referred to as “detection fibers”) at right angle to the “excitation fiber” at every sensor region. The second fiber provides the time delay necessary to temporally resolve the signals originating at the sensor regions as seen in Figure 1 (for details see [36, 37]). This allows for an increased spatial resolution in multisensor arrays, as we demonstrated with an array of 100 crossed fiber junctions in a space of 6cm×6cm [38]. This work is part of our effort to develop sensing elements (i.e., individual sensor junctions) for this sensor-array architecture; the long-term goal is to incorporate various sensing elements for different sensing tasks into this architecture.

Optical oxygen sensors are commonly based on the ability of molecular oxygen to quench the luminescence of certain transition metal complexes [39] or porphyrin derivatives [40]. The variation of the luminescence signal, 𝐼(and the luminescence decay time 𝜏) with the dissolved oxygen concentration DO, is described by the Stern-Volmer equations [41]𝐼0𝐼=𝜏0𝜏=1+𝐾SVDO,(1) where 𝐼0 and 𝜏0 are the respective luminescence intensity and luminescence lifetime in the absence of oxygen. The Stern-Volmer quenching constant 𝐾SV=𝑘𝑞𝜏0 depends on the diffusion-dependent bimolecular quenching constant 𝑘𝑞, which in turn is given by 𝑘𝑞=4𝜋𝑎𝐷(103𝑁𝐴). 𝐷 is the sum of the diffusion coefficients of oxygen and luminophores, 𝑎 is the sum of the molecular radii, and 103𝑁𝐴 is the number of molecules per millimole [42], where 𝑁𝐴 is Avogadro’s number.

The ruthenium complex, Ru[(Phen)3]Cl2, was chosen for this work as it exhibits high oxygen sensitivity because of its long unquenched lifetime 𝜏0 (~0.8 μs in ethanolic solution). Furthermore, it absorbs strongly at blue-green wavelengths and has a relatively large Stokes shift. As seen from (1), oxygen pressure may be measured by monitoring luminescence intensity or luminescence lifetime. We have adopted both approaches to DO determination.

3. Experimental

3.1. Reagent and Materials

The complex tris(1,10-phenanthroline)ruthenium(II) chloride hydrate (hereafter referred to as “RuPhen”), poly(ethylene glycol) diacrylate (PEG-DA) with an average molecular weight of 575 and, poly(ethylene glycol) dimethacrylate (PEG-MA) with a molecular weight of 1000, 2,2-dimethoxy-2-phenylacetophenone (DMPA), 3-(trichlorosilyl) propyl methacrylate (TPM), trimethylolpropane triacrylate (TPT), sulfuric acid, and hydrogen peroxide (30% V/V in water) were obtained from Sigma Aldrich Chemical Company (Milwaukee, WI). The laser dye Rhodamine 110 (Rh110) was purchased from Exciton Corporation (Dayton, OH). Deionized water (DI) with a resistivity of 18 MΩ·cm was used for all aqueous experiments. All reagents were used as received without further purification.

3.2. Chemical Modification of the Fiber Core

All setups used multimode silica fibers with a TECS (trademark of 3 M Corp.) cladding (FT-200-UMT with core diameter of 200 μm) and SMA 905 connectors purchased from ThorLabs, Inc. (Newton, NJ). The fibers had core and cladding refractive indices of 𝑛co=1.457 and 𝑛cl=1.404, respectively. The sensor regions themselves were prepared by first removing the fiber jackets of both fibers comprising the fiber junction by mechanical stripping. Subsequently, the cladding layers were removed by rubbing with a cotton swab immersed in acetone and the exposed core was cleaned using Piranha solution (3 : 1 sulfuric acid : hydrogen peroxide) to eliminate any organic residue. The first step in the synthesis of the new cladding was treating the silica core of the fiber with 1 M nitric acid solution to activate the hydroxyl groups of the silica. Then, a silanation reaction was carried out by exposing the fiber core surface with a 1 mM solution of TPM in a nitrogen atmosphere, at 18°C for 1 hr [4345]. Once the reaction was completed, the fiber was removed from the reaction cell, washed with DI water and ethanol, and dried in a nitrogen stream. The chlorosilane modification of the core is necessary to create gels that adhere covalently to the fiber surface.

3.3. Sensor Preparation

For molecules physically entrapped in polymers, thin PEG-based films were synthesized by mixing 1 mL of PEG-DA 575 with 10 mg of DMPA to prepare the PEG-DA 575 precursor solution, which was then diluted with water to make a 40% V/V solution (referred as coating solution). Separate solutions of 7.0 nM RuPhen complex (i.e., the oxygen sensor molecule) and 1.3 nM Rh110 (the intensity reference molecule) were made in ethanol. Each luminophore solution was added separately to the coating solution to make a 1 : 10 (V/V) mixture (designated as doped-coating solution). Highly crosslinked PEG-DA (TPT-PEG-DA 575) precursor solution was synthesized by adding, while stirring, 10 mg of DMPA and 300 μL of TPT to 1 mL of PEG-DA 575 and further sonicated for 10 min. The coating solution was formulated by diluting this mixture with water to make a 40% V/V solution. Subsequently the luminophore solution (either RuPhen or Rh110) was mixed with the 40% (V/V) coating solution to prepare a final 1 : 10 (V/V) doped-coating mixture. Additional PEG films were prepared using PEG-MA with a molecular weight of 1000 g/mol. Iin this case, first the waxy solid was heated at 37°C to melt the polymer and then 1 mL of the liquefied reagent was used to prepare the precursor solution.

After preparing the doped-coating mixtures, the fibers were placed in the flow cell such that orthogonal fiber-fiber junctions were created (see Figure 1) with the TPM-modified fiber cores just touching. The first sensor region in the array (i.e., the region that is excited first by the laser pulse) contains the intensity reference molecule Rh110. The RuPhen oxygen sensor is located in the second fiber-fiber sensor junction. 2.0 μL of each doped-coating mixture were placed in the corresponding region using an automatic pipette and the gel formation was initiated upon UV-exposure (365 nm) of the doped-coating mixture [4650]. The free-radical copolymerization of methacrylate and terminal acrylate groups between the PEG derivatives (of the doped-coating mixture) and the TPM functionality (bound to the fiber-core) creates polymer films covalently attached to the fiber core and where the luminophore is physically entrapped in the gel matrix. A Photon Technology International (PTI) 150-Watt Xenon lamp (Birmingham, NJ) with a narrow bandpass interference filter (J43-057—Edmund Optics) was used to cure PEG-gel films in approximately 1 minute. After gel formation the flow cell was filled with deionized water.

3.4. Instrumentation

A typical setup used in these experiments is shown in Figure 1. A dye laser PTI PL201 (using the laser dye Coumarin 460—Exciton Corporation) pumped by a nitrogen laser PTI PL2300 (pulse width 0.6 ns, pulse energy 1.4 mJ, pulse repetition rates of 3–10 Hz) served as the excitation source for the sensor array system. A pair of lenses was employed to couple the light into the excitation fiber. The sensors were mounted in a home-built flow cell filled with deionized water. Oxygen and nitrogen mixtures from pure gas cylinders (Praxair grades 2.6 and 4.8, resp.) were used to achieve the range of DO concentrations by bubbling these gas mixtures into the flow cell. The gases were premixed in a separate mixing chamber and delivered through 1/4-in.-i.d. flexible PVC tubing into the flow cell.

The luminescence emitted by the sensor molecules was captured by the fibers crossing at right angle to the fiber carrying the excitation light pulse and guided to the detector. A narrow bandpass filter (610 nm Edmunds J43-079) was used in front of the Burle C31034A photomultiplier tube (PMT1; with Peltier cooling) to select the RuPhen luminescence and the Rh110 reference pulses. As seen in Figure 1, the array is designed such that the region containing the reference is excited first; also, the lengths of the detection fibers are chosen such that the emission of the Rh110 reference arrives first at PMT1. The RuPhen luminescence pulse is delayed by passage through an additional stretch of fiber (9 m) to arrive at PMT1 after the reference signal has subsided. Thus, the output current of PMT1, which was analyzed with a LeCroy LC564DL digital storage oscilloscope (DSO) with a bandwidth of 1 GHz and sampling rate of 4 GS/s, displays both region signals sequentially in one trace. A trigger for the DSO was generated using a second PMT2 (RCA 1P28, 1.6-ns rise time), which collected light scattered from the front of the excitation fiber. A LabVIEW (National Instruments) program was written to collect and analyze the data. All the experiments presented here were repeated from three to four times, and the peak ratios were measured and averaged. The RuPhen referenced intensity 𝐼(𝐼RuPhen/𝐼Rh110) is calculated as the luminescence integrated intensity of RuPhen itself divided by the luminescence integrated intensity of the reference dye.

4. Results and Discussion

4.1. Polymer Selection and Indicator Leaching Test

For oxygen sensing in aqueous media, the polymer matrix must have high oxygen permeability to allow the rapid interaction of the dissolved oxygen with the indicator entrapped in that matrix, must be of mechanical and chemical stability, and must be optically clear. It is also worth noting that the low oxygen solubility in water at ambient conditions and the lower DO diffusion coefficient compared to the one in gas phase result in a poor sensitivity and response to DO. For PEG-based gels it has been demonstrated that their mesh size can be increased by the introduction of water to the precursor solution, which reduces the degree of polymer cross-linking [51]. An increase in porosity allows for faster diffusion of analytes through the gel; however, indicator leaching is increased. Leaching is an important aspect in the preparation of water-compatible sensors. The performance, repeatability, and the response of such sensors are associated to the degree of leaching. Several factors will determine the extension of the leak; with respect to the indicator molecules their molecular size, their solubility in both water and polymer matrices have to be considered for the fabrication of aqueous-phase sensors. With regard to the matrix polymer its chosen porosity is a compromise between indicator-molecule leaching from the solid matrix to the aqueous medium and analyte diffusion from aqueous environment to the entrapped molecule. Another important restriction in the selection of a polymer matrix is its level of hydrophobicity. Based on the molecular size of Rh110 and RuPhen the best matrix candidates are PEG-DA with a molecular weight of 575 or/and PEG-MA with a molecular weight of 1000. The higher the polymer molecular weight is, the more hydrophobic and porous it becomes.

Using these polymers in the design of aqueous sensors, we tested the degree of indicator leaching associated with them. For this, the change in luminescent intensities of the two regions placed in a water-filled flow cell was monitored every 10 minutes for 2 days. Figure 2 curves (a) and (b) show the integrated referenced RuPhen intensity as function of time for sensors based on PEG-DA 575 and PEG-MA 1000 polymer films, respectively. The observation of significantly larger leaching from the PEG-MA 1000 film suggests an increase in indicator-molecule leaching as the matrix molecular weight increases. This can be understood in term of two factors: first, it appears that PEG-MA 1000 is not sufficiently hydrophobic to retain RuPhen in its matrix and to avoid its diffusion to the aqueous medium. The second factor is the larger porosity obtained with PEG-MA 1000: the larger the mesh size the faster is the indicator-molecule diffusion toward the aqueous medium. Even though leaching is considerably less for a PEG-DA 575 matrix, the loss of 71% of RuPhen is quite detrimental to sensor performance. An alternative for decreasing indicator leaching is the incorporation of alkyl groups in the polymer structure [13, 5254]: this provides not only high porosity (because of the bulky groups) but also hydrophobicity. Therefore, we modified the PEG-DA 575 polymer by adding a hydrophobic crosslinker, TPT. The incorporation of this trifunctional monomer to the PEG-DA 575 precursor mixture improves the resistance to indicator leaching by more than 20%, as shown in Figure 2 curve (c). Even though an intensity loss of 50% is quite large, we still can detect sufficient signal intensity from the retained sensor molecules. Increasing the TPT concentration further results in an increase in the coating-solution refractive index, which will lead to refractive losses at the junctions and eventually to the loss of guiding condition of the fiber. After the initial leaching ceased, the TPT-PEG-DA 575-based DO sensor shows excellent stability, which is described in the following sections. While for this sensor one goal was to find a PEG formulation that would provide physical entrapment of the dye which minimized leaching but not overly compromised sensor response time, covalent attachment of the sensor dye to the PEG matrix is an alternative approach to control leaching, which we employed in other work [55].

4.2. Quenching Sensitivity

The sensor sensitivity can be expressed by the overall quenching efficiency (𝐸DO) given by 𝐸DO=𝐼N2𝐼O2𝐼N2×100%,(2) where 𝐼N2 and 𝐼O2 represent the referenced RuPhen intensities in fully deoxygenated and fully oxygenated water, respectively.

The oxygen concentrations in air-saturated and oxygen-saturated solutions were calculated to be 8.5 and 99.1 ppm at 22°C, respectively. This calculation was based on the solubility equation of oxygen in water [56]. For each DO concentration, 1000 luminescence pulses were averaged. All measurements were performed at room temperature. Figure 3 shows the response of the TPT-PEG-DA 575-based oxygen sensor to different DO concentrations.

Applying (2) to the data shown in Figure 3, we obtained a value of 𝐸DO=84%. This shows that, because of the issues discussed in the Introduction, the quenching response of the oxygen sensor is reduced in the aqueous phase compared to the gas-phase value of 𝐸DO=98% (found in our laboratory [57]). However, the measured quenching efficiency still constitutes a sufficiently high overall quenching response to attain good sensor resolution, especially at low oxygen concentrations. The DO-sensor described by Xu et al. [9] used Ru(II)-tris(4,7-diphenyl-1,10-phenanthroline) as sensor molecule enclosed in a sol-gel matrix and ratiometric detection exhibited a DO quenching efficiency of ~92%. The sensor studied by Koo et al. [13], which employed platinum (II) porphyrins derivatives incorporated in organically modified silicate nanoparticles, attained a value of 𝐸DO=97%. Our sensing system has a somewhat lower quenching performance, but the sensors mentioned above are for use in confined areas such cells, where our sensor, which has the potential for remote monitoring, is not applicable.

4.3. Sensor Reversibility

We used the cycle-purging method to check the reversibility of TPT-PEG-DA 575 films with RuPhen physically entrapped in this highly crosslinked matrix. The referenced sensor intensity (𝐼) is plotted in Figure 4 for varying oxygen concentrations. Upon switching from nitrogen-saturated to oxygen-saturated solution, 𝐼 decreased with increasing DO concentration. This process was repeated multiples times with no significant change in 𝐼 between the initial nitrogen-saturated solution, indicating complete reversibility of the sensor going from high to low DO concentration and vice versa. In addition, this sensor showed at least a 98% recovery each time after changing the environment, implying a good reproducibility.

Over the range of oxygen concentrations that we studied and for the measurement duration, we did not observe significant photobleaching of the Ru complex, which in turn may indicate singlet oxygen created in the quenching process can rapidly diffuse away before causing oxidative damage, as was reported in [58].

4.4. Sensor Response Time

The sensor response time was tested by alternating oxygenation/deoxygenation cycles, which is shown in Figure 5. The calculated transition time is of the order of 5–7 seconds; however we believe that the sensor itself has a shorter response time because the measured response time includes the time required to change between oxygenated and deoxygenated environments in the flow cell and the attached tubing. Therefore the reported value of the response time is an upper limit of the actual one. The thickness of the PEG film plays an important role in the sensor response time, particularly in the highly crosslinked films used here, which could reduce analyte diffusion to the sensor molecules and negatively impact the sensor response time. The crosslinking density in the PEG films used here still allows for effective analyte diffusion into the sensing volume between the fibers; the total film thickness through which the analyte must travel to reach the sensing volume is estimated to be less than 1 mm. Should faster response times be required, the creation of highly porous files using a microtemplating approach in combination with covalent attachment of the sensor dye would offer an alternative [55, 59].

4.5. Sensor Calibration and Linearity

The RuPhen luminescence quenching by oxygen is described by the Stern-Volmer equations (1), which are used to estimate the linear range of our sensor. Figure 6 shows Stern-Volmer plots of the RuPhen referenced intensity and luminescence lifetime versus DO concentration for the TPT-PEG-DA 575-based sensor; the dotted line represents the best linear fit generated by the least-squares method. The sensor has a good linearity over the whole range of DO concentrations tested, which may indicate negligible matrix heterogeneity effects.

The good 𝐸DO and linearity in the Stern-Volmer plot may result from the structure and hydrophobic nature of the film used in the sensor preparation. The introduction of TPT may provide good hydrophobicity and sufficient porosity to create multiple channels, allowing oxygen to travel efficiently towards the sensor molecule entrapped in the matrix. Because oxygen has higher affinity for hydrophobic media, this could lead to an improvement in sensitivity as well as in linearity.

The linear fit of Figure 6 yields a Stern-Volmer quenching constant of 𝐾SV=0.048ppm1. Using the unquenched lifetime 𝜏0=850ns (measured in the highly crosslinked film submerged in water) we obtain for the bimolecular quenching constant the value of 𝑘𝑞=1.8×109 L mol−1s−1, which is in the same order of magnitude to the value reported for oxygen sensing with RuPhen in solution (𝑘𝑞=4.2×109 L mol−1s−1 [60]).

To calculate the pressure resolution afforded by the referenced sensor, we use the sensor sensitivity defined as Δ𝑑𝐼𝐼(DO)=𝑑(DO)ΔDO,(3) where Δ𝐼(DO) is measurement uncertainty of the referenced-integrated intensities at a given dissolved oxygen concentration, and ΔDO is the corresponding resolution of the DO concentration. Using for the referenced-intensity uncertainty the 95% confidence band 𝑡×𝜎𝑁(𝐼), that is, the standard deviation of the mean 𝜎𝑁(𝐼) at a given DO concentration with the appropriate t-value for 𝑛=3, we obtain for oxygen concentrations of 3.91 ppm and 48.98 ppm, a sensor resolutions of 0.78 ppm and 0.82 ppm, respectively. To calculate the DO detection limit, DO(min,𝐼) afforded by the sensor, we use the sensitivity at DO=0, which using (1) is given by 𝑑𝐼/𝑑(DO)|DO=0=𝐾SV𝜏0 With 𝜎𝑁(𝐼0)=90.83ppm, we obtain for the detection limit DO(min,𝐼)=179ppb. The same calculation for the lifetime data yields for oxygen concentrations of 3.91 ppm and 48.98 ppm a sensor resolution of 0.41 ppm and 0.32 ppm, respectively, and a detection limit of 140 ppb. Thus, in spite of the two evanescent coupling processes occurring in our sensor platform, the detection limit for DO is within one order of magnitude of the detection limit reported for commercially available extrinsic fiber sensors, where the sensor dyes are mounted at the distal end of a fiber and interrogated with the phase-shift method [10]. McDonagh and coworkers [54] reported a sol-gel-based optical DO sensor with a resolution of 0.18 ppm and a detection limit of 5 ppb suggesting that sol-gel films are better candidates for DO sensing than the modified-PEG-DA gels. We investigated the use of sol-gel matrices in our fiber sensors, but while the sensor-molecule encapsulation was successful, the sol-gel (TEOS-based gels) was not covalently attached to the fiber core, which caused a fluctuating sensor response because of (slight) respective movements of the fibers constituting a fiber-fiber sensor junction.

4.6. Interferences

As explained above, the oxygen sensing mechanism is based on the quenching of the RuPhen luminescence, which, however, may also be quenched by other species. The interference because of anions, cations, and some oxidizing agents was tested by first purging the sensor chamber with nitrogen to create at oxygen-free atmosphere. Subsequently each interfering solution was injected into the flow cell while continually monitoring the referenced RuPhen intensity in an oxygen free aqueous medium. The overall sensor quenching responses to the dissolved species are depicted in Figure 7. It can be seen that the sensor response is affected by the presence of F and Fe+3. However, the sensor response to these interferences is much smaller than the sensor response to oxygen. The matrix thus protects the entrapped indicator from the environment, preventing interference with the luminescent properties of the sensor molecules. Without this shielding of the indicator, its luminescence would behave unpredictably in a given media, making calibration even of ratiometric luminophores difficult or impossible.

In addition to the interference studies, a DO sensor is most likely to be applied in biological media (e.g., blood, rivers, food processing, etc.) where the contamination of the sensor surface with proteins and cells could affect the diffusion of oxygen into the sensor. As a consequence, the antibiofouling properties of the TPT-PEG-DA 575 films were examined. The effects of nonspecific protein binding to the PEG-DA-film surface were investigated by the addition of bovine serum albumin (BSA, 0.3–5% (W/V) aqueous solution) to the flow-cell containing the fiber-fiber junctions immersed in DI water (see Figure 7 right-lower corner). First, the flow cell was purged with nitrogen, then, a 0.3% (W/V) BSA solution was added and allowed to interact with the fiber regions in the aqueous medium for 20 minutes. After this time period, oxygen was introduced to the flow cell with a DO concentration of 1.5 ppm. As expected the RuPhen intensity was quenched; after ~75 minutes the sensor was flushed with nitrogen and the RuPhen intensity recovered to the initials values, indicating that no protein binding on the sensor surface occurred. We observed the same trends with a BSA concentration of 5% (W/V) with the sensor maintaining its recovery time and sensitivity. The results from the nonspecific protein-binding assay show the importance of encapsulating molecules inside the protective PEG-DA matrix. The antibiofouling property, biocompatibility, and nontoxicity of poly(ethylene glycol) have been reported by different authors [6165], making this material suitable for biological applications. The vulnerability of the sensor to metal ions, notorious oxidizing species, and biological materials was prevented, demonstrating the capability of this sensor for environmental DO monitoring. The sensor performance towards DO response was also tested over time and the long-term quenching stability of DO sensor was established over a period of 6 months when the fiber-sensor array was continuously immersed in water.

5. Conclusions

Dissolved-oxygen sensing was implemented in a crossed-fiber sensor configuration, with sensing based on luminescence quenching of ruthenium complexes entrapped in a PEG-DA matrix. In spite of the weak evanescent coupling of the two fibers in a sensor junction, both lifetime- and intensity-based oxygen measurements exhibited good repeatability and signal-to-noise ratio with a limit of detection of ~0.8 ppm. With the use of second crossed-fiber junction to measure the excitation pulse intensity and referencing the ruthenium sensor signal, the intensity-based and lifetime-based measurements were of similar quality with regard to resolution and detection limit. Chemical modification of the PEG-DA matrices was shown to improve the performance of the dissolved-oxygen sensor. By employing organic precursors, which impart surface hydrophobicity, the quenching response can be comparable to that obtained in the gas phase. The PEG-DA matrices have sufficient density to prevent movement of the two fibers with respect to each other, while still allowing fast oxygen diffusion to the encapsulated ruthenium sensor for fast response times. Thus, the results indicate the potential for remote dissolved-oxygen sensing for use in a wide range of environments.

Acknowledgments

This work was supported by a grant from the University of Wisconsin Groundwater Research Program. Additional support through a University of Wisconsin-Milwaukee Dissertation Fellowship and a Dr. and Mrs. George Sosnovsky Award for Graduate Research (MVR) are gratefully acknowledged. The authors would like to thank Paul Henning for assistance with writing of the data acquisition software.