A Simple and Facile Glucose Biosensor Based on Prussian Blue Modified Graphite String

Table 1

Comparison of the analytical performance characteristics of the string-based glucose biosensor with other Prussian blue modified electrode.

ElectrodePB modification methodSensitivity
(µA·mM−1·cm−2)
Linear range
(mM)
LOD
(µM)
Response time (sec)References

Chi/GOx/PB-G
string electrode
Dipping641.30.03~1.010<3This work
GOx/Chi-PB/Chi-Au electrodeElectrodeposition0.002~0.40.4<10[10]
GO-GCE/PB/GOx/ChitosanElectropolymerization408.70.005~1.20.12<5[11]
GOx/PB-FCNF/GCEChemical process and casting35.940.02~12505[12]
GOx/Ferri-COs-
SPCEs
Drop-coating0.677
(µA·mM−1)
~33.31380<20[13]

PB: Prussian blue; GOx: glucose oxidase; Chi: chitosan; GO: graphene oxide; GCE: glassy carbon electrode; PB-FCNF: Prussian blue nanostructures/carboxylic group-functionalized carbon nanofiber nanocomposites; Ferri-COs: hexacyanoferrate (III) mixed with chitosan oligomers; SPCE: screen printed carbon electrode.