Research Article

Geometry Optimization Approaches of Inductively Coupled Printed Spiral Coils for Remote Powering of Implantable Biomedical Sensors

Table 2

Optimal inductive link coil designs by using Iterative Procedure.

ParametersPrimary coilSecondary coil

Outer diameter )80 (mm)20 (mm)
Inner diameter ()10 (mm)11 (mm)
Width of conductor ()2.5 (mm)0.3 (mm)
Spacing between conductors ()7.5 (mm)0.6 (mm)
Number of turns ()45
at  MHz78%