Research Article
Geometry Optimization Approaches of Inductively Coupled Printed Spiral Coils for Remote Powering of Implantable Biomedical Sensors
Table 2
Optimal inductive link coil designs by using Iterative Procedure.
| Parameters | Primary coil | Secondary coil |
| Outer diameter ) | 80 (mm) | 20 (mm) | Inner diameter () | 10 (mm) | 11 (mm) | Width of conductor () | 2.5 (mm) | 0.3 (mm) | Spacing between conductors () | 7.5 (mm) | 0.6 (mm) | Number of turns () | 4 | 5 | at MHz | 78% |
|
|