Selection of MEMS Accelerometers for Tilt Measurements

1Institute of Micromechanics and Photonics, Faculty of Mechatronics, Warsaw University of Technology, ul. Boboli 8, 02-525 Warsaw, Poland
2Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology, Technická 2896/2, 616 69 Brno, Czech Republic

Correspondence should be addressed to Sergiusz Łuczak; lp.ude.wp.rthcm@kazcul.s

Received 23 December 2016; Accepted 5 February 2017; Published 30 March 2017

Academic Editor: Pietro Siciliano

Copyright © 2017 Sergiusz Łuczak et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

In order to build a tilt sensor having a desired sensitivity and measuring range, one should select an appropriate type, orientation, and initial position of an accelerometer. Various cases of tilt measurements are considered: determining exclusively pitch, axial tilt, or both pitch and roll, where Cartesian components of the gravity acceleration are measured by means of low-g uni-, bi-, tri-, or multiaxial micromachined accelerometers. 15 different orientations of such accelerometers are distinguished (each illustrated with respective graphics) and related to the relevant mathematical formulas. Results of the performed experimental study revealed inherent misalignments of the sensitive axes of micromachined accelerometers as large as 1°. Some of the proposed orientations make it possible to avoid a necessity of using the most misaligned pairs of the sensitive axes; some increase the accuracy of tilt measurements by activating all the sensitive axes or reducing the effects of anisotropic properties of micromachined triaxial accelerometers; other orientations make it possible to reduce a necessary number of the sensitive axes at full measurement range. An increase of accuracy while using multiaxial accelerometers is discussed. Practical guidelines for an optimal selection of a particular micromachined accelerometer for a specific case of tilt measurement are provided.

1. Introduction

Micromachined accelerometers belonging to Microelectromechanical Systems (MEMS) are employed in many kinds of various devices [1], owing to their small dimensions (currently even less than 2 mm), easy integration with electronics, high shock-resistance, high reliability, satisfactory accuracy, low power consumption, and low cost. They realize measurements of various types of acceleration (whether constant or variable), its derivatives (jerk, jounce) and integrals (velocity, position). New unexpected applications emerge constantly in, for example, smartwatches, underwater equipment like scuba diving computers (e.g., Vyper Air and Novo, Cobra 3, DX, D6i Novo by Suunto Company), new generation of safe motorcycle ABS [2] or traction control systems (e.g., Motorcycle Stability Control by Bosch Corp. [3]), or novel designs of commonly known objects, like the electronic gaming die presented in [4]. One of the most strategic fields of application is various types of navigation systems, proposed, for example, in [5, 6].

Besides typical measurements of acceleration related to vibration or motion, another basic application of accelerometers is tilt measurements [7]. Such measurements are employed in many devices, usually for the purposes of positioning, aligning, leveling, or navigating. However, these measurements are correct only under static or quasi-static conditions. Moreover, in the case of quasi-static conditions, especially at considerable frequencies, additional errors due to amplitude attenuation [8] must be taken into account.

As far as specific applications of tilt measurement are concerned, accelerometers are embedded in such equipment as various portable digital devices (e.g., cellular phones, photo cameras, computer projectors, GPS receivers, and game consoles). They can be applied directly in small devices like mobile microrobots [9] (especially when they are a part of the only navigation system available, like in the case of some in-pipe or biomimetic robots, proposed, e.g., in [10, 11]), or indirectly, as embedded in smartphones, as reported, e.g., in [12]. One can also find a lot of untypical applications of MEMS accelerometers used for tilt measurements related to life sciences, for example, investigation of movements of animals [13] or humans [1416] and rehabilitation equipment, for example, exoskeletons presented in [1719], to name only a few.

Due to the fact that different mathematical formulas can be employed and different types of MEMS accelerometers are available, tilt measurements can be realized in various ways. Besides, orientation of the accelerometer is yet another important issue comprehensively addressed in the paper.

In order to perform a tilt measurement, the most advantageous solution is usually to apply a triaxial MEMS accelerometer and use arc tangent function [20]. Nevertheless, there are cases, specified in the paper, when other types of MEMS accelerometers are more preferable than triaxial sensors, for example, due to their inherent imperfections. Reasons for employing mathematical formulas other than arc tangent function are minutely discussed in [20]. For instance, if the accuracy of simple tilt measurements is not satisfactory, usually complex solutions are searched for, based, for example, on data fusion. Yet, sometimes a satisfactory increase of accuracy can be obtained by arranging an appropriate orientation of the MEMS accelerometer.

Despite common application of MEMS accelerometers, it is hard to find any satisfactory information on the aforementioned problems in the related publications. Therefore, we decided to address the issue of what type of MEMS accelerometer (uni-, bi-,tri-, or multiaxial) should be used in order to meet requirements related to certain case of tilt measurement, defining also its orientation and initial position as well as analyzing the respective sensitivity resulting from application of a given mathematical formula.

Owing to the guidelines presented in the paper, on the one hand, it is possible to build the simplest tilt measurement unit based on MEMS accelerometers, being at the same time the most cost-effective yet ensuring a satisfactory accuracy, and, on the other hand, to obtain possibly high accuracy at the cost of complicating the measurement.

2. Component Angles of Tilt

The most common way of expressing tilt is to determine its two component angles pitch and roll (defined with accordance to aeronautical nomenclature, as, for example, in [21]) illustrated in Figure 1 against the components of the gravitational acceleration . Often, it is the auxiliary angle that is defined as the roll, for example, in [2228], as in some cases such approach is more convenient, since accelerometers respond directly to this angle and not to angle .

Figure 1: Components of the tilt and the gravitational acceleration.

Arbitrary tilt angle (axial tilt) included between acceleration and axis has been resolved into two component angles (pitch) and (roll) or alternatively auxiliary angle . Angles and are contained within vertical planes and , (the planes are generally not orthogonal with respect to one another), that is, angles between horizontal plane and axis or , respectively. The coordinate system is immobile (its axis is vertical), whereas the coordinate system xyz is fixed to the mobile tilt sensor located at its origin. Angles and are interdependent and their values result from the value of angle . In order to illustrate angle, line k has been created as a result of the intersection of vertical plane and horizontal plane .

Angle between axis and is contained within a tilted plane ; it is the angle by which the tilt sensor rotates around axis x, while tilted in two phases (the first phase is rotation by pitch around axis y while it still overlaps axis ). It should be noted that most of the related formulas presented later in the text are analogous for angle and . Relation between the considered angles is as follows [21, 29]:

However, determining roll on the basis of (1) usually results in a decrease of the related accuracy [29] (see (21)).

From this point on, angles , , , and will be designated with additional subscripts, depending on the type of the considered tilt measurement (single- or dual-axis, as illustrated in Figure 2, or any of these two) and the formula employed for their determination, as specified in Nomenclature. The subscripts have been introduced in order to allow us to unequivocally distinguish between various cases introduced later in the text.

Figure 2: (a) Single-axis tilt sensor; (b) dual-axis tilt sensor.

It was accepted that, in case of single-axis tilt measurements, the only component tilt angle that appears is pitch .

3. Mathematical Formulas

The formulas presented in this section can be found in numerous related publications, for example, [20, 2235]. As it results from Figure 1, the component tilt angles can be calculated basically as follows:

However, the vertical component acceleration allows us to use the following formulas:and additionally,

If tilt is to be measured over one axis only, (5)–(8) get simplified, as one of the components of acceleration ( or ) will be then of zero value (e.g., (8) becomes identical to (9)). So, the equations will get transformed as follows:

While determining the arbitrary tilt angle , which in fact is an axial tilt defined in [33] and may be regarded as a pseudo dual-axis tilt measurement, the following relations are valid:

If or , then the measurement is actually a single-axis tilt measurement, where is determined instead of .

Application of (14)–(16) may be interesting in such measurements of axial tilt as, for example, directional drilling [36] or while monitoring an object against losing its balance, since pitch and roll are inconvenient to be used in such cases.

Additionally, as discussed in Section 8, under certain conditions, measurement of axial tilt may be more advantageous than measurement of pure pitch, as far as the resultant accuracy is concerned.

Besides the presented formulas, there are still other ways of determining the tilt angles. As suggested in [24], it is advantageous in some cases to combine an arc sine formula ((3) or (4), (15)) with an arc cosine formula ((5) or (6) and (10) or (11), (14)) in a simple way or as a weighted average [30, 37].

4. Orientations of the Accelerometer

Possibilities of applying certain accelerometer in particular tilt measurement are illustrated in Table 1.

Table 1: Tilt sensors versus types of accelerometer.

Formulas for determining pitch and roll given in Table 1 are listed for each case in the order of the highest measurement sensitivity (see Section 5); colors of the sensitive axes of the accelerometers, as they appear in the presented figures, agree with the accepted convention (see Nomenclature).

It should be added that obtaining a full measurement range of the component tilt angles is possible by observing the sign of the vertical component acceleration , which is positive over the domain of , while negative over the remaining angular range (thus, at least two sensitive axes of the accelerometer are required) [31].

Additional comments to the cases specified in Table 1:(i)Cases  1-A, 2-A, 2-B, 5-A, and 5-B are commonly applied and are well known.(ii)Case  1-B represents a pseudo dual-axis tilt (axial tilt) referred to in [33].(iii)Case  2-A, typically used for single-axis tilt measurement, can be accepted for dual-axis tilt measurement using an untypical orientation illustrated in cases 3-B and 4-B, which are an alternative to case 2-B; the resulting advantage is obtaining the full measurement range for both pitch and roll, using only 2 sensitive axes. Choice between 3-B and 4-B is related to the sensitivity of the measurements: at 3-B the highest sensitivity is obtained for = 0° and = 90°, whereas at 4-B for = 90° and = 0°; the sensitivity itself is variable, just as in case 2-B.(iv)Cases  5-A and 6-A are not recommended if it is striven for obtaining a higher sensitivity of the measurement, since none of the accelerometer sensitive axes should overlap axis while tilted, just as in case 7-A. However, additional calculations may be then necessary to convert angle into . Analogously, the same applies accordingly to cases 8-A and 9-A.(v)For cases  6-A, 6-B, and 7-B, see Section 7 for explanation.(vi)Cases  8-A, 8-B, and 9-A refer to a sensor presented in [38], which has been used as an example of a multiaxial accelerometer. Application of such sensor makes it possible to increase the sensitivity of tilt measurements even by 13% [30].

While using a multiaxial accelerometer, before calculating tilt angles, Cartesian components of the gravitational acceleration must be determined first. With regard to the accelerometer presented in cases 8-A, 8-B, and 9-A, the following formulas are true [38]:where , , , and are projections of the vector of the gravitational acceleration onto the sensitive axes of the accelerometer, which are inclined by 54.7° (colors of the sensitive axes, represented by the bold arrows, denote their numbers with accordance to the accepted nomenclature).

Except for selecting an appropriate case of tilt measurement (as specified in Table 1), it is also important to set the initial position of the accelerometer in an advantageous way. The relevant data related to the cases illustrated in Table 1 are included in Table 2.

Table 2: Initial positions of the accelerometers.

5. Sensitivity of Tilt Measurements

Even though the nominal values of the tilt angles calculated according to the presented equations are the same, yet they are determined with different sensitivity, thus different accuracy. So, application of particular formula significantly affects the accuracy of the measurement.

5.1. Sensitivity Related to Angles , , and

The appropriate formulas can be derived from (3)–(8), as demonstrated in [20]. They refer to pitch in the same measure as to roll (while expressed by the auxiliary angle ), so a tilt angle has been introduced. A relation between the sensitivity (expressed in rad−1) and the tilt angle can be defined as [24]

Likewise, for tilt angles and we can obtain [20]

Sensitivities and , related to the arc sine and the arc cosine formulas, decrease down to zero at a tilt angle of 90° or 0°, respectively, what is very disadvantageous. However, application of the arc tangent formulas ((7)-(8), (12)-(13), and (16)) ensures a constant and at the same time the highest value of the sensitivity.

As far as the axial tilt is concerned, sensitivity refers to , refers to , and refers to .

5.2. Sensitivity Related to Angle

On the basis of (1), the related sensitivity can be determined as

The sensitivity is expressed on the plot in g/rad; thus, it is variable over the range of ca. m/s2rad. It should be noted that not every combination of angles and is possible, since the angles are interdependent.

In the case of (9), the related sensitivity can be expressed by the following formula:which is also variable over the range of m/s2rad, this time exactly as . Both formulas are strongly nonlinear. However, sensitivity is at least equal to (often higher). Plot of (20) is illustrated in Figure 3.

Figure 3: Variability of the sensitivity expressed by (20).

In numerous applications, accuracy of tilt measurements (here represented by the sensitivity) is an important issue (even though there are some trivial cases when this problem is of no concern). Then, the most convenient mathematical relationship should be used, taking into account, first of all, the required accuracy and then simplicity of the related computations (thus, the resultant speed and cost of the measurement unit). Moreover, it must not be overlooked that the higher the accuracy of the measurements the higher the number of the sensitive axes of the accelerometer.

To conclude, it can be stated that in most of the cases the best solution is to use arc tangent formulas, that is, (7)–(9), (12)-(13), and (16), since they feature the highest sensitivity, provided that the number of the available sensitive axes is sufficient. However, there are also cases when the other formulas are preferred, as minutely discussed in [20].

It is worth mentioning that the sensitive axes need not to be orthogonal. Other spatial configurations are acceptable, and sometimes even more advantageous. Then, higher sensitivity can be obtained over some angular regions at the cost of its lower value over the other regions. Another disadvantage is the fact that the related computations are then more complicated, as briefly discussed in [30].

6. Accelerometer Selection

The required number of the sensitive axes results from type of tilt measurement, the measurement range, and the employed mathematical formula. Type of the selected accelerometer must comply with this number (the number of its sensitive axes may be alternatively higher).

While 3 sensitive axes are required, instead of employing 1 triaxial or 1 multiaxial accelerometer, it may be considered to apply 3 uniaxial or 2 biaxial accelerometers, or 1 uniaxial and 1 biaxial accelerometer. Such approach is more laborious and expensive; however it provides an opportunity to build a more accurate tilt sensor. This concept is accepted by such companies as MicroStrain Inc. or Sentera Technology Corp., whose sensors have the sensitive axes of the constituent accelerometers precisely calibrated and aligned (or have the relevant misalignments compensated for) [39].

However, even though some time ago triaxial MEMS accelerometers were not commercially available, nowadays it is quite the opposite: some manufacturers offer only triaxial devices, having abandoned production of biaxial or uniaxial versions. Yet, sometimes it is worthwhile to apply uniaxial or biaxial MEMS accelerometers. For instance, the alignment error of the sensitive axes of biaxial accelerometers is sometimes very small, that is, of 0.01° [28], unlike in the case of triaxial accelerometers, whose z-axis may be poorly aligned and additionally feature worse accuracy due to higher noise, what results from their mechanical structure determined by the fabrication process (this shortcoming is evident especially in the case of surface micromachining), which in fact is semi-three-dimensional [40].

Characteristics of particular mathematical formulas with respect to the sensitivity, the number of the required sensitive axes, and the measuring range of the applied accelerometers as well as complication of the related computation algorithm (proposed, e.g., in [31]) are minutely discussed in [20]. A brief summary is presented in Table 3.

Table 3: Characteristics of the considered mathematical formulas.

One more problem to be considered is the operational parameters of the accelerometers. Two basic issues are the bandwidth and the supply voltage. If a possibly high accuracy is pursued, the supply voltage should be stabilized and of the highest value, whereas the bandwidth should be as small as possible, for example, 0-1 Hz. Then, the noise of the accelerometer is the lowest [28].

7. Imperfections of MEMS Accelerometers

Despite many aforementioned advantages of MEMS accelerometers, they feature some imperfections, of which the most significant, as far as tilt measurements are concerned, are thermal and long-term drifts of the output signals (both of the offset and the scale factor), misalignments (connected also with a lack of a physical datum surface and nonorthogonality of the sensitive axes), and cross-axis sensitivities. As far as thermal drifts are concerned, even some commercial sensors have this problem solved; for example, 16201 [41] of the ADIS series by Analog Devices Inc. Misalignments are dealt with rather seldom, for example, in the accelerometer described in [39]. However, referring to the cross-axis sensitivities, many models compensating for this errors have been proposed so far, for example, in [32, 36]. Less significant shortcomings are necessity of calibrating the accelerometer by the user (in the case of some accelerometers, custom equipment, like the one presented in [42], may be then required), anisotropy of the mechanical structure, sensitivity to electrostatic discharges, and thermal, mechanical, and electrical hysteresis.

However, if a basic type of accelerometer is used (with no compensation for the aforementioned shortcomings, yet much cheaper), and at the same time a relatively high accuracy of tilt measurements must be obtained, especially the thermal errors should be dealt with, as proposed, for example, in [22].

With regard to anisotropy, in some cases this imperfection can be eliminated. For instance, if a single triaxial accelerometer is used in a tilt measurement that does not require using all its output signals, the sensitive axis with the worst performance should be inactive. As a result from the relevant catalog data, the signal related to vertical axis of MEMS accelerometers (with respect to their packaging surface) is usually more noisy, compared to the horizontal axes and . So, it is more advantageous in single-axis tilt measurements to use the orientation of the accelerometer like in case 6-A (Table 1) instead of case 5-A.

As far as dual-axis measurements are concerned, two options are reasonable. If pitch is to be determined with a higher sensitivity, orientation illustrated in case 6-B should be used. On the other hand, if it is roll that is more significant, the accelerometer should be oriented as in case 7-B.

With regard to the mechanical and electrical hysteresis it should be mentioned that its influence can be essentially reduced by subjecting the accelerometer to excited vibration (quasi-static conditions of operation). Moreover, as it results from our own experimental studies, this phenomenon is of minor significance in the case of MEMS accelerometers [43].

It must not be forgotten that each measurement case requires a precise alignment of the applied MEMS accelerometer, or appropriate compensation for the alignment errors, as addressed in [44]; otherwise a considerable loss of accuracy of determining the tilt must be taken into account.

As already mentioned, some of the listed imperfections are dealt with by manufacturers, who offer more sophisticated versions of the accelerometers, some even adjusted for tilt measurements, like ADIS 16201 by Analog Devices Inc. [41] or SCA100T by Murata Electronics Oy, called even an inclinometer [45].

8. Experimental Studies

In order to determine the expected increase of accuracy related to untypical orientations presented in Table 1, appropriate experimental studies were performed, in which we used the test rigs described in [42, 43] and two types of triaxial MEMS accelerometers ADXL 330 and 327 (two pieces of each type) by Analog Devices Inc.

First, the experiments revealed that the sensitive axes of the tested accelerometers were not perpendicular, with inherent mutual misalignments in the range of 0.05°–1° (much higher than specified in the relevant datasheets). Moreover, values of the misalignments were different for each pair of axes and for each of the tested accelerometers. Such large value of the inherent misalignments of the sensitive axes of MEMS accelerometers is a very important issue in the case, when the misalignments are not compensated for in a numerical way (it is not possible to physically align the sensitive axes with respect to one another). As proven in [44], this kind of misalignment may directly increase the error of tilt measurement.

However, taking into account values of the inherent misalignments, it is possible to choose a more convenient orientation of the accelerometer, for example, case 6-A instead of 5-A, case 6-B or 7-B instead of 5-A. The result may be a decrease of the measurement error even as large as 1°, as our experiments proved.

By the way, it is not necessary to know values of the inherent misalignments. In order to choose a better orientation, a simple calibration should be performed for each orientation, and on the basis of the obtained results, the more accurate orientation should be accepted.

Due to the aforementioned considerable inherent misalignments of the sensitive axes of the tested accelerometers it was not possible to evaluate accuracy increase related to single-axis measurements while accepting orientation specified in case 7-A as a better alternative to cases 5-A and 6-A (see Table 1). For the same reason, at this stage of experiments, it was not possible to verify how significant is the anisotropy of the tested triaxial MEMS accelerometers, and whether the related effects could be slightly reduced (applying case 6-A instead of 5-A; see Table 1).

Finally, in order to verify if it is more advantageous to use simplified formulas (10)–(13) instead of (5)–(8) in cases 5-A and 6-A, respectively, other series of experiments were performed. However, no significant difference was observed (relative changes of respective maximum errors as well as sums of errors were found to be no bigger than 0.4%). Nevertheless, in case of a bigger misalignment of the inactive sensitive axis (e.g., of ca. 1°), increase of accuracy may be considerable, as proved in [44].

Further experimental study will be performed in order to evaluate increase of accuracy of tilt measurement obtained owing to employing orientations specified in cases 6-A, 7-A, 6-B, and 7-B (see Table 1). Nevertheless, at this stage of experimental works (where the inherent misalignments have not been compensated for), it can be stated that the experimental results proved that it is worthwhile to reduce the inherent misalignments by choosing appropriate orientation of the accelerometer.

9. Conclusions

A practical approach to determining tilt with various types of MEMS accelerometers has been presented. The accepted idea of using components of the gravity acceleration for this purpose allows us to build a tilt sensor with miniature overall dimensions while employing MEMS accelerometers.

Results of our own experimental studies, reported, for example, in [24, 31], prove that using such accelerometers it is possible to detect tilt angles with accuracy of ca. few tenths of degree arc. However, our further experimental works reported in [4244] revealed considerable imperfections of such accelerometers: the inherent misalignments between the sensitive axes (reaching up to 1°) as well as anisotropic metrological properties of the sensitive axes. Therefore, we proposed few original configurations of the applied MEMS accelerometer as one of the ways of reducing effects of the observed imperfections.

While arranging the discussed tilt sensor, the following issues must be considered (in the respective order):(i)Type of tilt measurement (single/dual-axis)(ii)Type of the applied accelerometers (uni-, bi-, tri-, or multiaxial; measurement range of the accelerometer should be ≥1 g, yet possibly small at the same time)(iii)Sensitivity of tilt measurement (determined mostly by the employed mathematical formula and noise of the accelerometer)(iv)Simplicity of mathematical processing of the output signals

Then, appropriate measurement case presented in Table 1 can be selected along with a corresponding initial position of the applied accelerometer, as specified in Table 2.

It must be realized that the best solution often does not exist. Usually, a trade-off must be sought for. Nevertheless, the paper shows how to build an optimal tilt sensing device. If typical arrangements, represented by cases 1-A, 2-A, 2-B, 5-A, and 5-B (see Table 1), do not satisfy certain requirements, other alternatives are proposed.

When a higher accuracy must be obtained, the following original orientations can be employed:(i)Cases 6-A, 6-B, and 7-B (see Table 1) make it possible to reduce the effects of anisotropic metrological properties of triaxial MEMS accelerometers; the accuracy can be increased even by 1°.(ii)cases 7-A and 9-A (see Table 1) make it possible to employ all the sensitive axes while realizing single-axis tilt measurements; the accuracy can be slightly increased; however, misalignments of the sensitive axes must be reduced to at least ca. 0.05° (whether physically or by compensation); otherwise, the accuracy may be even lower (as experimentally confirmed).(iii)A multiaxial accelerometer may be applied, as illustrated in cases 8-A, 8-B, and 9-A (see Table 1), to increase the accuracy (e.g., by 13%).

On the contrary, if for some reason (e.g., simplicity or speed of computations) the number of the sensitive axes should be reduced in a dual-axis tilt measurement over the full measurement range, cases 3-B and 4-B (see Table 1) may be accepted.

Still another issue is tilt measurement over a small range. At present, it is rather a difficult task using MEMS accelerometers, since only few models feature a small measuring range. Moreover, the range is relatively large anyway (e.g., 0.5 g in the case of SCA100T by Murata Electronics Oy, MXA6500EP or MXR2999EL by MEMSIC Inc.). However, it may be expected that smaller measuring ranges will be available soon, since some laboratory works in this field have already been started, as reported, for example, in [46].

It should be also mentioned that, in order to improve the accuracy, the output signals generated by MEMS accelerometers can be processed in a more sophisticated way, as proposed, for example, in [47]. Another way of achieving this goal is a sophisticated procedure of calibration of the accelerometers, as reported, for example, in [48, 49].

Nomenclature

:Gravitational acceleration
, , :Cartesian components of acceleration
:Projection of acceleration measured in th sensitive axis of a multiaxial accelerometer
:th sensitive axis of multiaxial accelerometer (1st, dark blue; 2nd, green; 3rd, yellow; 4th, red; observed in all the figures)
:Pitch
:Auxiliary tilt angle
:Roll
:Axial tilt (arbitrarily oriented tilt angle)
:Either pitch or angle
:Sensitivity
:Either , , , , or
:Quantity “” related to arc sine function
:Quantity “” related to arc cosine function
:Quantity “” related to arc tangent function
:Quantity “” related to either single- or dual-axis tilt measurement
:Quantity “” related to single-axis tilt measurement
:Quantity “” related to dual-axis tilt measurement
case -:Orientation defined in Table 1, Row , Column .

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Authors’ Contributions

Sergiusz Łuczak designed and wrote the manuscript; Robert Grepl analyzed sensitivity of determining roll angle and created the related graph; Maciej Bodnicki revised the manuscript; all the other figures were created by the acknowledged persons.

Acknowledgments

The authors would like to thank graduates of the Faculty of Mechatronics, Warsaw University of Technology, for creating the high-quality 3D graphics: Ms. Katarzyna Guzowska, B.S. Eng. (also graduate of the Academy of Fine Arts) (Figure 2 and figures contained in Table 1), Mr. Grzegorz Ekwiński, M.S. Eng. (Figure 1), Mr. Wojciech Załuski, M.S. Eng. (Figure 1), and Mr. Marcin Karczewski, M.S. Eng. (figures contained in Table 1).

References

  1. N. Abbate, A. Basile, C. Brigante, A. Faulisi, and F. La Rosa, “Modern breakthrough technologies enable new applications based on IMU systems,” Journal of Sensors, vol. 2011, Article ID 707498, 7 pages, 2011. View at Publisher · View at Google Scholar · View at Scopus
  2. S. Łuczak, “Tilt measurements in BMW motorcycles,” in Recent Global Research and Education: Technological Challenges, R. Jabloński, Ed., pp. 287–293, Springer, Cham, Switzerland, 2017. View at Google Scholar
  3. Active Safety Systems. Motorcycle Stability Control (MSC) Enhanced, Robert Bosch GmbH, Stuttgart, Germany, 2015.
  4. E. MacDonald, R. Salas, D. Espalin et al., “3D printing for the rapid prototyping of structural electronics,” IEEE Access, vol. 2, pp. 234–242, 2014. View at Publisher · View at Google Scholar · View at Scopus
  5. N. Navidi, R. J. Landry, J. Cheng, and D. Gingras, “A New Technique for Integrating MEMS-Based Low-Cost IMU and GPS in Vehicular Navigation,” Journal of Sensors, vol. 2016, Article ID 5365983, 16 pages, 2016. View at Publisher · View at Google Scholar · View at Scopus
  6. A. Ali and N. El-Sheimy, “Low-cost MEMS-based pedestrian navigation technique for GPS-denied areas,” Journal of Sensors, vol. 2013, Article ID 197090, 10 pages, 2013. View at Publisher · View at Google Scholar · View at Scopus
  7. J. S. Wilson, Sensor Technology Handbook, Newnes, Burlington, Mass, USA, 2005.
  8. A. Albarbar, A. Badri, J. K. Sinha, and A. Starr, “Performance evaluation of MEMS accelerometers,” Measurement: Journal of the International Measurement Confederation, vol. 42, no. 5, pp. 790–795, 2009. View at Publisher · View at Google Scholar · View at Scopus
  9. S. Fatikow and U. Rembold, Microsystem Technology and Microrobotics, Springer, Heidelberg, Germany, 1997. View at Publisher · View at Google Scholar
  10. M. Bodnicki and M. Sęklewski, “Design of small-outline robot—Simulator of gait of an amphibian,” in Recent Advances in Mechatronics, R. Jabłoński, M. Turkowski, and R. Szewczyk, Eds., pp. 77–81, Springer, Heidelberg, Germany, 2007. View at Google Scholar
  11. M. Bodnicki and D. Kamiński, “In-pipe microrobot driven by SMA elements,” in Mechatronics 2013. Recent Technological and Scientific Advances, T. Brezina and R. Jabloński, Eds., pp. 527–533, Springer, Cham, Switzerland, 2014. View at Google Scholar
  12. S. Pongnumkul, P. Chaovalit, and N. Surasvadi, “Applications of smartphone-based sensors in agriculture: a systematic review of research,” Journal of Sensors, vol. 2015, Article ID 195308, 18 pages, 2015. View at Publisher · View at Google Scholar · View at Scopus
  13. T. C. Nelson, P. Doukakis, S. T. Lindley et al., “Research tools to investigate movements, migrations, and life history of sturgeons (Acipenseridae), with an emphasis on marine-oriented populations,” PLoS ONE, vol. 8, no. 8, Article ID e71552, 2013. View at Publisher · View at Google Scholar · View at Scopus
  14. G. Rescio, A. Leone, and P. Siciliano, “Supervised expert system for wearable MEMS accelerometer-based fall detector,” Journal of Sensors, vol. 2013, Article ID 254629, 11 pages, 2013. View at Publisher · View at Google Scholar · View at Scopus
  15. H. Zhang, J. Zhang, D. Zhou et al., “Axis-exchanged compensation and gait parameters analysis for high accuracy indoor pedestrian dead reckoning,” Journal of Sensors, vol. 2015, Article ID 915837, 13 pages, 2015. View at Publisher · View at Google Scholar · View at Scopus
  16. C. S. Morère, Ł. Surazyński, A. R. Pérez-Tabernero, E. Vihriälä, and T. Myllylä, “MEMS technology sensors as a more advantageous technique for measuring foot plantar pressure and balance in humans,” Journal of Sensors, vol. 2016, Article ID 6590252, 9 pages, 2016. View at Publisher · View at Google Scholar · View at Scopus
  17. K. Bagiński, D. Jasińska-Choromańska, and J. Wierciak, “Modelling and simulation of a system for verticalization and aiding the motion of individuals suffering from paresis of the lower limbs,” Bulletin of the Polish Academy of Sciences: Technical Sciences, vol. 61, no. 4, pp. 919–928, 2013. View at Publisher · View at Google Scholar · View at Scopus
  18. B. G. Loh and J. Rosen, “Kinematic analysis of 7 degrees of freedom upper-limb exoskeleton robot with tilted shoulder abduction,” International Journal of Precision Engineering and Manufacturing, vol. 14, no. 1, pp. 69–76, 2013. View at Publisher · View at Google Scholar · View at Scopus
  19. D. Jasińska-Choromańska, K. Szykiedans, J. Wierciak et al., “Mechatronic system for verticalization and aiding the motion of the disabled,” Bulletin of the Polish Academy of Sciences: Technical Sciences, vol. 61, no. 2, pp. 419–431, 2013. View at Publisher · View at Google Scholar · View at Scopus
  20. S. Łuczak, “Guidelines for tilt measurements realized by MEMS accelerometers,” International Journal of Precision Engineering and Manufacturing, vol. 15, no. 3, pp. 489–496, 2014. View at Publisher · View at Google Scholar · View at Scopus
  21. S. Popowski, “Determining pitch and roll in inexpensive land navigation systems,” Journal Aeronautica Integra, vol. 1, pp. 93–97, 2008 (Polish). View at Google Scholar
  22. M. Horton and C. Kitchin, “Dual-axis tilt sensor based on micromachined accelerometers,” Sensors, vol. 13, no. 4, pp. 91–94, 1996. View at Google Scholar · View at Scopus
  23. I. Frosio, F. Pedersini, and N. A. Borghese, “Autocalibration of triaxial MEMS accelerometers with automatic sensor model selection,” IEEE Sensors Journal, vol. 12, no. 6, pp. 2100–2108, 2012. View at Publisher · View at Google Scholar · View at Scopus
  24. S. Łuczak, W. Oleksiuk, and M. Bodnicki, “Sensing tilt with MEMS accelerometers,” IEEE Sensors Journal, vol. 6, no. 6, pp. 1669–1675, 2006. View at Publisher · View at Google Scholar · View at Scopus
  25. R. Dao, Inclination Sensing with Thermal Accelerometers, MEMSIC, 2002.
  26. STMicroelectronics, Tilt Measurement Using a Low-G 3-Axis Accelerometer, STMicroelectronics, 2010.
  27. Kionix, Tilt Sensing with Kionix MEMS Accelerometers, Kionix, Ithaca, NY, USA, 2005.
  28. “Low Cost ±2 g Dual Axis Accelerometer With Duty Cycle Output, ADXL 202E,” Analog Devices Inc., 2000.
  29. S. Łuczak, “Single-axis tilt measurements realized by means of MEMS accelerometers,” Engineering Mechanics, vol. 18, pp. 341–351, 2011. View at Google Scholar
  30. S. Łuczak and W. Oleksiuk, “Increasing accuracy of tilt measurements,” Engineering Mechanics, vol. 14, pp. 143–154, 2007. View at Google Scholar
  31. S. Łuczak, “Advanced algorithm for measuring tilt with MEMS accelerometers,” in Recent Advances in Mechatronics, R. Jabłoński, M. Turkowski, and R. Szewczyk, Eds., pp. 511–515, Springer, Heidelberg, Germany, 2007. View at Google Scholar
  32. M. Šipoš, P. Pačes, J. Roháč, and P. Nováček, “Analyses of triaxial accelerometer calibration algorithms,” IEEE Sensors Journal, vol. 12, no. 5, pp. 1157–1165, 2012. View at Publisher · View at Google Scholar · View at Scopus
  33. S. Łuczak, “Specific measurements of tilt with MEMS accelerometers,” in Mechatronics. Recent Technological and Scientific Advances, R. Jabloński and T. Brezina, Eds., pp. 705–711, Springer, Berlin, Germany, 2011. View at Google Scholar
  34. D. Jurman, M. Jankovec, R. Kamnik, and M. Topič, “Calibration and data fusion solution for the miniature attitude and heading reference system,” Sensors and Actuators, A: Physical, vol. 138, no. 2, pp. 411–420, 2007. View at Publisher · View at Google Scholar · View at Scopus
  35. M. Carminati, G. Ferrari, R. Grassetti, and M. Sampietro, “Real-time data fusion and MEMS sensors fault detection in an aircraft emergency attitude unit based on kalman filtering,” IEEE Sensors Journal, vol. 12, no. 10, pp. 2984–2992, 2012. View at Publisher · View at Google Scholar · View at Scopus
  36. W. Yang, B. Fang, Y. Y. Tang, J. Qian, X. Qin, and W. Yao, “A robust inclinometer system with accurate calibration of tilt and azimuth angles,” IEEE Sensors Journal, vol. 13, no. 6, pp. 2313–2321, 2013. View at Publisher · View at Google Scholar · View at Scopus
  37. R. Grepl, “Balancing wheeled robot: effective modelling, sensory processing and simplified control,” Engineering Mechanics, vol. 16, no. 2, pp. 141–154, 2009. View at Google Scholar
  38. S. Bütefisch, A. Schoft, and S. Büttgenbach, “Three-axes monolithic silicon low-g accelerometer,” Journal of Microelectromechanical Systems, vol. 9, no. 4, pp. 551–556, 2000. View at Publisher · View at Google Scholar · View at Scopus
  39. Sentera Technology, AX301 Three-Axis Accelerometer Module, Sentera Technology, 2003.
  40. V. Kaajakari, Practical MEMS, Small Gear Publishing, Las Vegas, Nev, USA, 2009.
  41. Analog Devices, Programmable Dual-Axis Inclinometer/Accelerometer ADIS 16201, Analog Devices, 2006.
  42. S. Łuczak, “Dual-axis test rig for MEMS tilt sensors,” Metrology and Measurement Systems, vol. 21, no. 2, pp. 351–362, 2014. View at Publisher · View at Google Scholar · View at Scopus
  43. S. Łuczak, “Experimental studies of hysteresis in MEMS accelerometers: a commentary,” IEEE Sensors Journal, vol. 15, no. 6, pp. 3492–3499, 2015. View at Publisher · View at Google Scholar · View at Scopus
  44. S. Łuczak, “Effects of misalignments of MEMS accelerometers in tilt measurements,” in Mechatronics 2013. Recent Technological and Scientific Advances, T. Brezina and R. Jabloński, Eds., pp. 393–400, Springer International Publishing, Cham, Switzerland, 2014. View at Google Scholar
  45. Murata Manufacturing Co., Ltd., MEMS Sensors & Sensing Elements, Murata Manufacturing Co., Ltd., 2016.
  46. D. Yamane, T. Konishi, T. Matsushima et al., “Sub-1G MEMS accelerometer,” in Proceedings of the 12th IEEE SENSORS 2013, Baltimore, Maryland, November 2013. View at Publisher · View at Google Scholar · View at Scopus
  47. W. T. Latt, K. C. Veluvolu, and W. T. Ang, “Drift-free position estimation of periodic or quasi-periodic motion using inertial sensors,” Sensors, vol. 11, no. 6, pp. 5931–5951, 2011. View at Publisher · View at Google Scholar · View at Scopus
  48. D. Lee, S. Lee, S. Park, and S. Ko, “Test and error parameter estimation for MEMS - based low cost IMU calibration,” International Journal of Precision Engineering and Manufacturing, vol. 12, no. 4, pp. 597–603, 2011. View at Publisher · View at Google Scholar · View at Scopus
  49. M.-S. Kim, S.-B. Yu, and K.-S. Lee, “Development of a high-precision calibration method for inertial measurement unit,” International Journal of Precision Engineering and Manufacturing, vol. 15, no. 3, pp. 567–575, 2014. View at Publisher · View at Google Scholar · View at Scopus