The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board

Table 1

Average reading of test strips measured by a blood glucose meter under different resistance conditions.

YSI reading 2.2 KΩ
(mg/dL)
2.4 KΩ
(mg/dL)
2.6 KΩ
(mg/dL)

38 mg/dL28.138.657.8
120 mg/dL91.9119.7155.9
220 mg/dL187.6220.0258.5
310 mg/dL281.5315.1338.5
478 mg/dL442472.8526.8