The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board

Table 2

Reading of test strips with different concentrations of standard solution obtained by a blood glucose meter under different resistance conditions, showing the influence of resistance on STD and CV.

YSI reading 2.2 KΩ
STD/CV
2.4 KΩ
STD/CV
2.6 KΩ
STD/CV

38 mg/dL3.7/—3.0/—5.8/—
120 mg/dL—/5.1—/3.7—/6.0
220 mg/dL—/5.3—/3.7—/5.0
310 mg/dL—/5.2—/3.5—/5.1
478 mg/dL—/4.9—/3.6—/5.1