The Influence of Blood Glucose Meter Resistance Variation on the Performance of a Biosensor with a Gold-Coated Circuit Board

Table 4

The range of test strip measurements under the resistances of 2.2 KΩ, 2.4 KΩ, and 2.6 KΩ for standard solutions with glucose concentrations of ≥100 mg/dL.

ResistanceWithin 15%Within 10%Within 5%

2.2 KΩ84/200 (42%)80/200 (40%)24/200 (12%)
2.4 KΩ200/200 (100%)200/200 (100%)169/200 (85%)
2.6 KΩ102/200 (51%)54/200 (27%)24/200 (12%)