Control solution test results from 3 lots of test strips with same electrode length but different electrode widths W (unit: mg/dL).

; Average75.4184234.1328.8412.2543.6
Standard deviation3.166.287.3810.738.5213.43
CV (%)4.193.413.153.262.072.47
YSI74185233326411548

; Average74.2184.8233.4327.2410.4546.3
Standard deviation2.293.445.207.655.338.43
CV (%)3.091.862.232.341.301.54