Control solution test results from 3 lots of test strips with same electrode length but different electrode widths W (unit: mg/dL).
| ; | Average | 75.4 | 184 | 234.1 | 328.8 | 412.2 | 543.6 | Standard deviation | 3.16 | 6.28 | 7.38 | 10.73 | 8.52 | 13.43 | CV (%) | 4.19 | 3.41 | 3.15 | 3.26 | 2.07 | 2.47 | YSI | 74 | 185 | 233 | 326 | 411 | 548 |
| ; | Average | 74.2 | 184.8 | 233.4 | 327.2 | 410.4 | 546.3 | Standard deviation | 2.29 | 3.44 | 5.20 | 7.65 | 5.33 | 8.43 | CV (%) | 3.09 | 1.86 | 2.23 | 2.34 | 1.30 | 1.54 | YSI | 74 | 185 | 233 | 326 | 411 | 548 |
|
|