Research Article
Design of Electrodes on Gold Test Strips for Enhanced Accuracy in Glucose Measurement
Table 4
Control solution test results from 3 lots of test strips for different electrode lengths but same electrode width (unit: mg/dL).
| ; | Average | 75.4 | 184 | 234.1 | 328.8 | 412.2 | 543.6 | Standard deviation | 3.16 | 6.28 | 7.38 | 10.73 | 8.52 | 13.43 | CV (%) | 4.19 | 3.41 | 3.15 | 3.26 | 2.07 | 2.47 | YSI | 74 | 185 | 233 | 326 | 411 | 548 |
| ; | Average | 72.8 | 183.8 | 238.9 | 333.9 | 415.4 | 557.6 | Standard deviation | 1.80 | 4.95 | 5.27 | 8.30 | 6.25 | 11.38 | CV (%) | 2.47 | 2.70 | 2.21 | 2.49 | 1.50 | 2.04 | YSI | 74 | 185 | 233 | 326 | 411 | 548 |
|
|