Review Article
Fundamentals, Recent Advances, and Future Challenges in Bioimpedance Devices for Healthcare Applications
Table 4
Comparison of bland Altman analysis results in studies of ICG-based cardiovascular parameters.
| Study | Sample size | Measurement | Reference method | ICG device | Differences Mean value | 95% CoIn |
| [242] | 42 patients (newborn - 16 yr) | CI (L/min/m2) | Doppler echocardiography | NiCaS, NI Medical, Israel | 0.05 | -0.77 to 0.87 |
| [243] | 25 healthy adults | CI (L/min/m2) | Magnetic resonance imaging (MRI) | FhysioFlow, Manatec Biomedical, France | 0.1 | -1.1 to 1.3 | 32 patients with chronic anemia | 1.5 | -0.6 to 3.6 |
| [244] | 11 healthy adults | SV (mL) | MRI | Cardiotronic, Osypka Medical, Germany | ~ 0 | -64 to 64 |
| [245] | 17 patients under hemodialysis | SV (mL) | Doppler echocardiography | NiCaS, NI Medical, Israel | 1.3 | -11.6 to 14.1 |
| [246] | 32 healthy adults | SV (mL) | Doppler echocardiography | non-commercial | -1.53 | -26.7 to 23.64 | CO (L/min) | 0.32 | -1.19 to 0.18 |
| [247] | 53 pregnant women | SV (mL) | Doppler echocardiography | ICG, Philips Medical Systems, USA | 2.5 | -15.7 to 20.7 | CO (L/min) | 0.2 | -1.3 to 1.7 |
|
|