Fundamentals, Recent Advances, and Future Challenges in Bioimpedance Devices for Healthcare Applications
Table 7
Bland Altman analysis results obtained in skin conductance studies.
| Study | Sample size | Measurement | Reference method | SCR device | Differences Mean value | 95% CoIn |
| [337] | 49 healthy elderly | Resting energy expenditure (kcal/24h) in the morning | Indirect calorimetry | Sensewear armband, Body- Media, USA | -166 | -438 to 105 | Resting energy expenditure (kcal/24h) at night | -187 | -427 to 53 |
| [338] | 133 type 2 diabetic patients and 41 control subjects | Sudomotor function (%) on hands | Quantitative sudomotor axon reflex test | EZSCAN, Impeto Medical, France | -4.3 | -23.5 to 15 | Sudomotor function (%) on foots | -1 | -14 to 12 |
|
|