Research Article

Differential Electroanalysis of Dopamine in the Presence of a Large Excess of Ascorbic Acid at a Nickel Oxide Nanoparticle-Modified Glassy Carbon Electrode

Table 5

Simultaneous electroanalytical determination of mixed DA-HCl and vitamin C real samples.

SampleAddedFoundaRecoveryb (%)
DA-HCL (μM)Vitamin C (mM)DA-HCL (μM)Vitamin C (mM)DA-HCLVitamin C

13.156.33.306.17
24.4610.54.3810.4
315.752115.7221