Research Article
Development of Capacitive Rain Gauge for Marine Environment
Table 1
Calibration data of rain gauge.
| Measuring point [ml] | Nominal voltage [V] | Measuring voltage [V] | Precipitation error [mm] | Uncertainty [mm] |
| 0 | 0 | 0.008 | 0.08 | 0.001 | 50 | 0.5 | 0.551 | 0.51 | 0.001 | 100 | 1.0 | 1.071 | 0.71 | 0.001 | 150 | 1.5 | 1.532 | 0.32 | 0.001 | 200 | 2.0 | 2.010 | 0.10 | 0.002 | 250 | 2.5 | 2.501 | 0.01 | 0.002 | 300 | 3.0 | 2.980 | -0.20 | 0.002 | 350 | 3.5 | 3.480 | -0.20 | 0.003 | 400 | 4.0 | 4.001 | 0.01 | 0.003 | 450 | 4.5 | 4.510 | 0.10 | 0.003 | 500 | 5.0 | 4.990 | 0.10 | 0.003 |
|
|