A Unified Calibration Paradigm for a Better Cuffless Blood Pressure Estimation with Modes of Elastic Tube and Vascular Elasticity

Table 6

Comparison results with prior work.

Calibration methodAcquired signals (measure location)SubjectsTest samples (pairs of BP)Accuracy w.r.t. cuff BP ( (MAD)/mmHg; )

LS [27]Only PPG: ear & toe200SBP: (7); /
DBP: (7); /
LS [8]ECG: hands & leg; PPG: finger; PCG: chest173SBP: (4.71); 0.95
DBP: (4.44); 0.84
LS [28]ECG: thorax; EIT: thorax166SPAP: (6); 0.87
PTP, three pairs [29]PPG: finger; IPG: wrist90SBP: (6); 0.88
DBP: (5); 0.88
oPTP [2]ECG: palms; PPG: finger169SBP: (4.4); 0.89
DBP: (4.6); 0.83
oPTP [30]ECG: arm; PPG: arm70SBP: (6); 0.80
DBP: no estimation
Multiple regression [31]ECG: back; PPG: thigh; BCG: thigh1147SBP: (5.50); 0.86
DBP: (3.86); 0.81
Probability distributions [32]Only ECG: chest3219SBP: (7.72); /
DBP: (9.45); /
aPTP, this work (MK-BH model)ECG: wrist & foot
PPG: finger
365SBP: (6.0); 0.83
DBP: (4.4); 0.74
aPTP, this work (dMK-BH model)ECG: wrist & foot
PPG: finger
365SBP: (5.3); 0.87
DBP: (4.0); 0.78

Note 1: “/” = not be estimated based on reported results or able to be reported in corresponding authors’ other work. Note 2: “†” = be approximately estimated from the corresponding Bland-Altman plots. Note 3: “‡” = the median value of the corresponding estimation accuracy index. Note 4: PCG = phonocardiogram; BCG = ballistocardiography; IPG = impedance plethysmography; EIT = electrical impedance tomography; SPAP = systolic pulmonary artery pressure.