Research Article

New Spectrophotometric and Conductometric Methods for Macrolide Antibiotics Determination in Pure and Pharmaceutical Dosage Forms Using Rose Bengal

Table 5

The intraday and interday precision and accuracy data for the studied drugs using the spectrophotometric method.

IntradayInterday
DrugTaken ( g mL−1)Found ( g mL−1)aRecovery %Precision RSD %bAccuracy Er %cTaken
(μg mL−1)
Found ( g mL−1)aRecovery %Precision RSD %bAccuracy Er %c

(I)5050.28100.560.290.565050.34100.680.390.68
(II)3030.33101.090.561.093030.17100.570.460.57
(III)1615.9999.950.19−0.051616.01100.070.250.07

aAverage of seven determinations.
bRSD %, percentage relative standard deviation.
cEr %, percentage relative error.