Different Spectrophotometric Methods Manipulating Ratio Spectra Applied for the Analysis of Aclidinium in Duaklir® Genuair® Inhalation Powder

Table 1

Regression and validation of data for determination of aclidinium by the proposed ratio derivative and ratio subtraction methods.

ParametersRatio derivativeRatio subtraction

Wavelength (nm)229238
Linearity range (μg/mL)5─50
(i) Regression equationya = bxb + aya = bxb + a
(ii) Slope (b)0.03800.0230
(iii) Intercept (a)−0.00070.0032
Coefficient of determination (r2)0.99970.9996
LOD (μg/mL)0.96811.1428
LOQ (μg/mL)2.93363.4630
Accuracy (%R)c101.55100.82
Precision (% RSD)cRepeatability1.2371.149
Intermediate precision1.7071.520

ya is the response of each method; xb is the concentration of aclidinium in μg/mL in each method; cvalues for 3 determinations of 3 different concentrations.