Research Article

Different Spectrophotometric Methods Manipulating Ratio Spectra Applied for the Analysis of Aclidinium in Duaklir® Genuair® Inhalation Powder

Table 4

Determination of aclidinium in Duaklir Genuair inhalation powder by the proposed ratio derivative, ratio subtraction, and the reported methods.

ParametersRatio derivativeRatio subtractionReported method [14]

na555
Average (% recovery)99.23100.2299.65
SD1.4621.5991.717
% RSD1.4741.5951.724
T (2.306)b0.9490.030
F (6.388)b1.4001.171

aNumber of measurements; bThe values in parenthesis are tabulated values of “t” and “F” at ().