Research Article

Low-Content Quantitation in Entecavir Tablets Using 1064 nm Raman Spectroscopy

Table 3

Performance of the regression models using raw Raman data of calibration set samples and validation set samples which were obtained by randomly divided the calibration samples.

PreprocessingRegionLVRMSECVR2RMSECRMSEP

Raw dataFull20.140.8960.090.28
1350–1700 cm−130.130.9250.080.30

BaselineFull30.140.8770.100.25
1350–1700 cm−130.150.9410.070.28

MSCFull30.160.8960.090.28
1350–1700 cm−140.140.9250.080.30

SNVFull30.130.9380.070.26
1350–1700 cm−120.130.9620.060.10

Baseline + MSCFull20.140.8960.090.28
1350–1700 cm−120.150.9250.080.30

Baseline+SNVFull30.140.9050.090.20
1350–1700 cm−130.130.9700.050.05

Baseline + D1 dataFull20.250.7970.120.33
1350–1700 cm−130.130.8920.090.25

Baseline + D2 dataFull30.160.8390.110.27
1350–1700 cm−140.180.8990.090.31