Research Article

Mathematically Processed UV Spectroscopic Method for Quantification of Chlorthalidone and Azelnidipine in Bulk and Formulation: Evaluation of Greenness and Whiteness

Table 1

Validation parameter results of the proposed spectroscopic methods for the simultaneous determination of CTL and AZE.

Validation parametersChlorthalidoneAzelnidipine
1DDRADR1DDRCS1DDRADR1DDRCS

Wavelength (nm)238.5211.8–267.2239.8226.9272.1328.4–286.1342.1257.3
Linearity range (μg/mL)1–151–151–151–151–151–151–151–15
Slope0.30251.09240.03670.06840.112916.4410.49860.0432
Intercept−0.0772−0.3106−0.0091−0.0063−0.01298.80350.1193−0.0089
Regression coefficient (r2)0.99960.99980.99980.99970.99990.99970.99970.9995
LOD (μg/mL)0.3050.2670.4080.3060.1280.3010.1860.328
LOQ (μg/mL)0.9240.7680.8060.9280.3870.9330.5990.995
Accuracy (mean % ± RSE)100.24 ± 0.9298.95 ± 1.0499.47 ± 1.3499.27 ± 1.6198.37 ± 1.4899.46 ± 0.54100.34 ± 1.49100.38 ± 1.41

Precision (%RSD)
Intraday1.0530.9451.4380.8931.2320.7151.7581.422
Interday1.1270.8460.8071.8031.4590.3970.9311.087

RSE: Relative standard error; RSD: Relative standard deviation.