Research Article

Simple Determination of Gemifloxacin Levels in Human Plasma using High-Performance Liquid Chromatography-Tandem Mass Spectrometry

Table 2

Stability of the proposed method for determining gemifloxacin levels in human plasma samples (n = 3).

Concentration (μg/mL)Short-term storage stabilityFreeze–thaw stabilityPostpreparative stability
% Accuracy% CV% Accuracy% CV% Accuracy% CV

0.0292.6613.43107.6310.54100.006.74
297.230.59101.826.72106.122.72

CV, coefficient of variation.